Seer Home System is an ambulatory electroencephalograph (EEG) system intended for at-home use for extended EEG and video EEG (vEEG) recordings and analysis. The Seer Home System includes the Sight Monitoring Hub, the Sense wearable, Monitoring Portal and associated accessory components (Disposable EEG Wires, WaterTabs, WaterTabs Application Kit, and Take-home Kit).

Published September 25, 2024 · Recall initiated: August 12, 2024

Summary

FDA records medical device Class II recall Z-3143-2024, published September 25, 2024 and initiated August 12, 2024. Recalling firm: SEER MEDICAL PTY LTD; Status: Ongoing; Reported quantity: 3,948 units; Lot: 1. Product Number: SM-03-8001-INT; UDI-DI: 09355751001037; Serial Numbers: 000089, 000090, 000091, 000092, 000095, 000182, 000201, 000207, 000208, 000211, 000225, 000245, 000258, 000259, 000285, 000286, 000289, 000290, 000291, 000292, 000296, 000297, 000298, 000299, 000301, 000305, 000307, 000308, 000309, 000310, 000312, 000313, 000316, 000321, 000322, 000323, 000094, 000183, 00028, 000302, 000303, 000306, 000311. 2. Product Number: SM-01-8001; UDI-DI: 09355751001020; Serial Numbers: 000074, 000084, 000113, 000118, 000119, 000187, 000188, 000191, 000223, 000224, 000225, 000226, 000228, 000229, 000230, 000232, 000239, 000241, 000242, 000243, 000244, 000245, 000247, 000250, 000251, 000259, 000262, 000263, 000277, 000279, 000280, 000282, 000283, 000286, 000287, 000288, 000289, 000290, 000291, 000292, 000294, 000297, 000299, 000300, 000301, 000302, 000303, 000304, 000305, 000306, 000307, 000308, 000309, 000310, 000312, 000313, 000315, 000317, 000318, 000319, 000320, 000321, 000322, 000323, 000324, 000327, 000328, 000329, 000333, 000335, 000336, 000337, 000339, 000340, 000341. 3. Product Number: Monitoring Portal; UDI-DI: 09355751001136; Serial Numbers: DMPD80RCMDFX, DMPFRRE7Q1GJ, DMPFT3PDQ1KQ, DMPFT43HQ1GG, DMPFT46JQ1KQ, DMPFT4CYQ1KQ, DMPFT4X1Q1KQ, DMPFV0HGQ1KQ, DMPFV178Q1KQ, DMPFV19SQ1KQ, DMPFV1AFQ1KQ, DMPFV1AXQ1KQ, DMPFV1DBQ1KQ, DMPFV1KQQ1KQ, DMPFV1L2Q1KQ, DMPFV1LNQ1KQ, DMPFV1PRQ1KQ, DMPFV1Q8Q1KQ, DMPFV1R2Q1KQ, DMPFV1R9Q1KQ, DMPFV1SRQ1KQ, DMPFV1TEQ1KQ, DMPFV1WYQ1KQ, DMPFV1YGQ1KQ, DMPFV1ZBQ1KQ, DMPFV1ZMQ1KQ, DMPFV20DQ1KQ, DMPFV20PQ1KQ, DMPFV219Q1KQ, DMPFV21GQ1KQ, F9FF60LYQ1KQ, F9FF70Z5Q1KQ, F9FF71PAQ1KQ, F9FFQ413Q1GJ, F9FFQC4WQ1GG, F9FFQF9VQ1GG. 4. Product Number: SM-01-9005; UDI-DI: 09355751001112; Serial Number: 006240129. 5. Product Number: SM-07-8001; UDI-DI: 09355751001129; Serial Numbers: 007230809, 007230810, 007230823, 007230825, SM-07-9001_230829, 007231004, 007231005, 007231012, 007231023, 007231025, 007231102, 007231108, 007231109, 007231115, 007231120, 007231124, 00720240129, 00720240205, 00720240207, 00720240314, 007240318, 007240319, 007240327, 007240325, 007240326, 007240402, 007240403, 007240507, 007240515, 007240516, 007240522, 007240527, 007240531, 007240612, 007240613, 007240624, 007240626, 007240627, 007240628, 007240702, 007240703. 6. Product Number: SM-01-9003-US; UDI-DI: 09355751001099; Serial Numbers: 230525, 004230609, 004230913, 004231121, 004240306, 004240403, 004240424, 004240502, 004240509, 004240531, 004240603, 004240626, 004240705. 7. Product Number: SM-01-9008; UDI-DI: 09355751001105; Serial Numbers: 00223058, 00223059, 00223066, 00224008, 00224011, 00224013, 00224017, 00224018, 00224021. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During CAPA investigations of sight production inconsistencies, it was highlighted that previous Electromagnetic Compatibility (EMC) assessments related to RF radiated emissions may not be reasonably equivalent to the current product due to design changes in the system. A new CAPA was opened to investigate the radiated emissions compliance issues. - Seer Home Systems distributed and retained in the field were not compliant with Electromagnetic Compatibility (EMC) testing according to IEC 60601-1-2 - Seer Home Systems, including accessories, were manufactured and distributed without application of proper design change and manufacturing controls - Seer Home Systems were distributed with labeling deficiencies

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
SEER MEDICAL PTY LTD
Distributor
Worldwide - US Nationwide distribution in the states of MN, NY, PA and the countries of Australia.
Lot
1. Product Number: SM-03-8001-INT; UDI-DI: 09355751001037; Serial Numbers: 000089, 000090, 000091, 000092, 000095, 000182, 000201, 000207, 000208, 000211, 000225, 000245, 000258, 000259, 000285, 000286, 000289, 000290, 000291, 000292, 000296, 000297, 000298, 000299, 000301, 000305, 000307, 000308, 000309, 000310, 000312, 000313, 000316, 000321, 000322, 000323, 000094, 000183, 00028, 000302, 000303, 000306, 000311. 2. Product Number: SM-01-8001; UDI-DI: 09355751001020; Serial Numbers: 000074, 000084, 000113, 000118, 000119, 000187, 000188, 000191, 000223, 000224, 000225, 000226, 000228, 000229, 000230, 000232, 000239, 000241, 000242, 000243, 000244, 000245, 000247, 000250, 000251, 000259, 000262, 000263, 000277, 000279, 000280, 000282, 000283, 000286, 000287, 000288, 000289, 000290, 000291, 000292, 000294, 000297, 000299, 000300, 000301, 000302, 000303, 000304, 000305, 000306, 000307, 000308, 000309, 000310, 000312, 000313, 000315, 000317, 000318, 000319, 000320, 000321, 000322, 000323, 000324, 000327, 000328, 000329, 000333, 000335, 000336, 000337, 000339, 000340, 000341. 3. Product Number: Monitoring Portal; UDI-DI: 09355751001136; Serial Numbers: DMPD80RCMDFX, DMPFRRE7Q1GJ, DMPFT3PDQ1KQ, DMPFT43HQ1GG, DMPFT46JQ1KQ, DMPFT4CYQ1KQ, DMPFT4X1Q1KQ, DMPFV0HGQ1KQ, DMPFV178Q1KQ, DMPFV19SQ1KQ, DMPFV1AFQ1KQ, DMPFV1AXQ1KQ, DMPFV1DBQ1KQ, DMPFV1KQQ1KQ, DMPFV1L2Q1KQ, DMPFV1LNQ1KQ, DMPFV1PRQ1KQ, DMPFV1Q8Q1KQ, DMPFV1R2Q1KQ, DMPFV1R9Q1KQ, DMPFV1SRQ1KQ, DMPFV1TEQ1KQ, DMPFV1WYQ1KQ, DMPFV1YGQ1KQ, DMPFV1ZBQ1KQ, DMPFV1ZMQ1KQ, DMPFV20DQ1KQ, DMPFV20PQ1KQ, DMPFV219Q1KQ, DMPFV21GQ1KQ, F9FF60LYQ1KQ, F9FF70Z5Q1KQ, F9FF71PAQ1KQ, F9FFQ413Q1GJ, F9FFQC4WQ1GG, F9FFQF9VQ1GG. 4. Product Number: SM-01-9005; UDI-DI: 09355751001112; Serial Number: 006240129. 5. Product Number: SM-07-8001; UDI-DI: 09355751001129; Serial Numbers: 007230809, 007230810, 007230823, 007230825, SM-07-9001_230829, 007231004, 007231005, 007231012, 007231023, 007231025, 007231102, 007231108, 007231109, 007231115, 007231120, 007231124, 00720240129, 00720240205, 00720240207, 00720240314, 007240318, 007240319, 007240327, 007240325, 007240326, 007240402, 007240403, 007240507, 007240515, 007240516, 007240522, 007240527, 007240531, 007240612, 007240613, 007240624, 007240626, 007240627, 007240628, 007240702, 007240703. 6. Product Number: SM-01-9003-US; UDI-DI: 09355751001099; Serial Numbers: 230525, 004230609, 004230913, 004231121, 004240306, 004240403, 004240424, 004240502, 004240509, 004240531, 004240603, 004240626, 004240705. 7. Product Number: SM-01-9008; UDI-DI: 09355751001105; Serial Numbers: 00223058, 00223059, 00223066, 00224008, 00224011, 00224013, 00224017, 00224018, 00224021.
Risk classification
Class II
Area (FDA)
device
Country
Australia

Related recalls

Questions about this recall

Was Seer Home System is an ambulatory electroencephalograph (EEG) system intended for at-home use for extended EEG and video EEG (vEEG) recordings and analysis. The Seer Home System includes the Sight Monitoring Hub, the Sense wearable, Monitoring Portal and associated accessory components (Disposable EEG Wires, WaterTabs, WaterTabs Application Kit, and Take-home Kit). recalled?

Yes. FDA published recall Z-3143-2024 on September 25, 2024. Reported hazard: During CAPA investigations of sight production inconsistencies, it was highlighted that previous Electromagnetic Compatibility (EMC) assessments related to RF radiated emissions may not be reasonably equivalent to the current product due to design changes in the system. A new CAPA was opened to investigate the radiated emissions compliance issues. - Seer Home Systems distributed and retained in the field were not compliant with Electromagnetic Compatibility (EMC) testing according to IEC 60601-1-2 - Seer Home Systems, including accessories, were manufactured and distributed without application of proper design change and manufacturing controls - Seer Home Systems were distributed with labeling deficiencies

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail