Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000
Published September 23, 2026 · Recall initiated: July 30, 2026
Summary
FDA records medical device Class II recall Z-3191-2026, published September 23, 2026 and initiated July 30, 2026. Recalling firm: Datex-Ohmeda, Inc.; Status: Ongoing; Reported quantity: 7 units; Lot: Serial Numbers: ANAS01357 ANAP00609 ANAQ00988 ANAS00586 ANAR00543 ANAP01088 ANAM00443. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Datex-Ohmeda, Inc.
- Distributor
- Worldwide - US Nationwide distribution.
- Lot
- Serial Numbers: ANAS01357 ANAP00609 ANAQ00988 ANAS00586 ANAR00543 ANAP01088 ANAM00443
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAEt Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen andSeptember 23, 2026
- FDAAisys CS2, Datex-Ohmeda anesthesia system, Part Number 1011-9055-000September 23, 2026
- FDAAisys, Datex-Ohmeda anesthesia system, Part Number 1011-9050-000September 23, 2026
- FDACaptus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154;September 23, 2026
- FDABoston Scientific Corporation StoneSmart Connect Console Contents: (1) Power Cord, (1) Power Adaptor, (1)DVISeptember 23, 2026
- FDAAisys. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatoryOctober 30, 2024
Questions about this recall
Was Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000 recalled?
Yes. FDA published recall Z-3191-2026 on September 23, 2026. Reported hazard: Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.
What should I do if I have this product?
Voluntary: Firm initiated · Letter