Philips EPIQ Elite Diagnostic Ultrasound System Model # 795234

Published September 30, 2026 · Recall initiated: August 7, 2026

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Due to software issue with ultrasound system, doppler sample is displayed shallower than its actual measurement location, potentially causing inaccurate blood flow velocity measurements and misdiagnosis.

What to do

Voluntary: Firm initiated

Recall details

Manufacturer
Philips Ultrasound, LLC
Distributor
Worldwide - US Nationwide distribution in the states of AL, CA, CT, FL, GA, KS, MD, MN, NJ, NY, OH, PA, SC, TN, TX, VA, WA, WI and the countries of Austria, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lithuania, Netherlands, Norway, Philippines, Poland, Romania, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan , United Kingdom, Vietnam.
Lot
All Philips EPIQ Elite Ultrasound Systems running software version 14.0, UDI(s): (01)00884838102514 (01)00884838107540 (01)00884838114159 (01)00884838124912 (01)00884838128125
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Philips EPIQ Elite Diagnostic Ultrasound System Model # 795234 recalled?

Yes. FDA published recall Z-3249-2026 on September 30, 2026. Reported hazard: Due to software issue with ultrasound system, doppler sample is displayed shallower than its actual measurement location, potentially causing inaccurate blood flow velocity measurements and misdiagnosis.

What should I do if I have this product?

Voluntary: Firm initiated

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Recalert also tracks expiration dates so you can sell stock before it becomes a loss.

Check your inventory (free)