Presource Packs Containing Stryker StrykeFlow PS and AHTO tube sets: 1) KIT, LAVH , Catalog Number: PB63VHSTC
Published October 7, 2026 · Recall initiated: August 18, 2026
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The kits contain tube sets with handpieces that may leak during use. The impacted devices can have the following failures: fluid leakage from the hand piece that can pool on the floor. The impact can range from user dissatisfaction to the potential for the user of slip and fall.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Cardinal Health 200, LLC
- Distributor
- US Nationwide distribution in the states of CA, IL, KY, MO, NC, NY, OH, TX.
- Lot
- 1) PB63VHSTC, UDI-DI: 10197106980866(each), 50197106980864(case), Lot Number: 25069FG2 ; 2) PB63VHSTC, UDI-DI: 10197106980866(each), 50197106980864(case), Lot Number: 25101FG2
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPresource Packs Containing Stryker StrykeFlow PS and AHTO tube sets: 1) KIT, LAP CHOLE , Cata…October 7, 2026
- FDAPresource Packs Containing Stryker StrykeFlow PS and AHTO tube sets: 1) KIT,GU DAVINCI , Cata…October 7, 2026
- FDAPresource Packs Containing Stryker StrykeFlow PS and AHTO tube sets: 1) KIT, GYN LAPAROSCOPY NEW HAN , Cata…October 7, 2026
- FDAOlympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.October 7, 2026
- FDAOlympus M3 Gold Autoclavable Lateral Telescope 70. Model Number: M3-70A.October 7, 2026
- FDAPALACOS R pro 40; PALACOS R pro 80. Model/Catalog Numbers: 5081286; 5081287. Radiopaque, poly (methyl meth…October 7, 2026
- FDAMedical convenience kits containing BD CliniClean Pericare Wipes within BD Urinary Catheter Trays ABDOMINA…October 7, 2026
- FDAPresource Convenience Kits: 1) KIT, PRE OP , Catalog Number: AMN548616A; 2) KIT, NICU AD…September 30, 2026
Questions about this recall
Was Presource Packs Containing Stryker StrykeFlow PS and AHTO tube sets: 1) KIT, LAVH , Catalog Number: PB63VHSTC recalled?
Yes. FDA published recall Z-3298-2026 on October 7, 2026. Reported hazard: The kits contain tube sets with handpieces that may leak during use. The impacted devices can have the following failures: fluid leakage from the hand piece that can pool on the floor. The impact can range from user dissatisfaction to the potential for the user of slip and fall.
What should I do if I have this product?
Voluntary: Firm initiated · Letter
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.
Recalert also tracks expiration dates so you can sell stock before it becomes a loss.