Siemens AXIOM Luminos dRF Max systems The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. The Axiom Luminos dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image.
Publicado el 8 de octubre de 2014 · Inicio del retiro: 2 de septiembre de 2014
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause lost images due to an automatic Radiology Information System (RIS) worklist update, thus causing examinations to be repeated. Image label may also be displayed incorrectly after images
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Siemens Medical Solutions USA, Inc
- Distribuidor
- Nationwide Distribution - US including UT, MO, CO, MA, MN, MS, OH, SC, MI, and NY.
- Lote
- Siemens AXIOM Luminos dRF Max system material #s: 10762471 with serial number: 5037.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
¿Vendes productos como este?