Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0009-2026

Allura Xper FD20/10; Model Numbers: 722029; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;

Publicado el 8 de octubre de 2025 · Inicio del retiro: 3 de septiembre de 2025

Resumen

FDA registra el retiro Z-0009-2026 de dispositivos médicos de Clase II, publicado el 8 de octubre de 2025 e iniciado el 3 de septiembre de 2025. Empresa responsable: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.; Estado: En curso; Cantidad reportada: 87 units (21 US, 66 OUS); Lote: Model Numbers: 722029; UDI-DI: 00884838054219; Serial Numbers: 71, 91, 7, 94, 62, 88, 36, 97, 21, 100, 30, 63, 60, 129, 95, 126, 110, 56, 55, 57, 31, 26, 49, 114, 128, 108, 103, 107, 81, 43, 52, 119, 118, 80, 78, 47, 106, 123, 96, 122, 14, 127, 101, 116, 102, 115, 19, 109, 120, 92, 69, 27, 46, 93, 51, 50, 124, 87, 20, 29, 32, 104, 34, 45, 59, 79, 24, 70, 39, 66, 82, 75, 40, 33, 23, 25, 85, 37, 44, 58, 38, 61, 98, 53, 89, 35, 48. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Distribuidor
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Guinea, Honduras, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Mauritius, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Norway, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, T®rkiye, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.
Lote
Model Numbers: 722029; UDI-DI: 00884838054219; Serial Numbers: 71, 91, 7, 94, 62, 88, 36, 97, 21, 100, 30, 63, 60, 129, 95, 126, 110, 56, 55, 57, 31, 26, 49, 114, 128, 108, 103, 107, 81, 43, 52, 119, 118, 80, 78, 47, 106, 123, 96, 122, 14, 127, 101, 116, 102, 115, 19, 109, 120, 92, 69, 27, 46, 93, 51, 50, 124, 87, 20, 29, 32, 104, 34, 45, 59, 79, 24, 70, 39, 66, 82, 75, 40, 33, 23, 25, 85, 37, 44, 58, 38, 61, 98, 53, 89, 35, 48;
Clasificación de riesgo
Class II
Área (FDA)
device
País
Netherlands

Retiros relacionados

Preguntas sobre este retiro

¿Allura Xper FD20/10; Model Numbers: 722029; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563; fue retirado del mercado?

Sí. FDA publicó el retiro Z-0009-2026 el 8 de octubre de 2025. Peligro reportado: The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter