Product Name: Nihon Kohden CNS-6200 Series Central Nurse Station, Model/Catalog Number: CN-6201 Software Version: software versions: 01-03, 01-04, 01-05, 01-06, 02-10, 02-11, and 02-40 Product Description: The device is intended for use by medical professionals to provide cardiac and vital signs monitoring for multiple patients within a medical facility. The CNS-6200 Series Central Nurse Station will display and record physiological data from up to forty telemetry receiver/transmitters and generates an alarm when a measured parameter falls outside a pre-set limit or when life threatening arrhythmia is detected. Arrhythmia detection and alarm determination are functions of the telemetry receivers/transmitters or individual bedside monitor. Component: No
Publicado el 14 de octubre de 2026 · Inicio del retiro: 8 de septiembre de 2026
Resumen
FDA registra el retiro Z-0089-2027 de dispositivos médicos de Clase II, publicado el 14 de octubre de 2026 e iniciado el 8 de septiembre de 2026. Empresa responsable: Nihon Kohden America LLC; Estado: En curso; Cantidad reportada: 808; Lote: Lot Code: CNS-6201 (PU-621RA), SW versions: 01-03, 01-04, 01-05, 01-06, 02-10, 02-11, and 02-40. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Due to a cybersecurity issue associate with a NULL Pointer Dereferences vulnerabiity. If exploited would potentially cause disruption of the centralized monitoring of multiple beds and the visualization of alarms.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Nihon Kohden America LLC
- Distribuidor
- US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN. KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NL, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, VT, QA, WA, WI, WV and WY.
- Lote
- Lot Code: CNS-6201 (PU-621RA), SW versions: 01-03, 01-04, 01-05, 01-06, 02-10, 02-11, and 02-40
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDABrand Name: MiniMed Flex Product Name: MiniMed Flex Insulin Pump Model/Catalog Numbers: MMT-8062; MMT-807214 de octubre de 2026
- FDAOnkos Surgical Inc. My3D Personalized Pelvic Implant, Porous, Left REF PV-ACREC-04M14 de octubre de 2026
- FDARollator, Bariatric, Basic Steel, Blue, mechanical walker with four legs and a metal frame intended for14 de octubre de 2026
- FDAOnkos Surgical Inc. My3D Personalized Pelvic Implant, Porous, Left REF PV-ACREC-04M14 de octubre de 2026
- FDARollator, Bariatric, Basic Steel, Burgundy, mechanical walker with four legs and a metal frame intended for14 de octubre de 2026
- FDADAVINCI TRUCUSTOM BUNDLE; Model/Item Number (Kit): 1. SAMM064-27, 2. SAMM064-28;14 de octubre de 2026
Preguntas sobre este retiro
¿Product Name: Nihon Kohden CNS-6200 Series Central Nurse Station, Model/Catalog Number: CN-6201 Software Version: software versions: 01-03, 01-04, 01-05, 01-06, 02-10, 02-11, and 02-40 Product Description: The device is intended for use by medical professionals to provide cardiac and vital signs monitoring for multiple patients within a medical facility. The CNS-6200 Series Central Nurse Station will display and record physiological data from up to forty telemetry receiver/transmitters and generates an alarm when a measured parameter falls outside a pre-set limit or when life threatening arrhythmia is detected. Arrhythmia detection and alarm determination are functions of the telemetry receivers/transmitters or individual bedside monitor. Component: No fue retirado del mercado?
Sí. FDA publicó el retiro Z-0089-2027 el 14 de octubre de 2026. Peligro reportado: Due to a cybersecurity issue associate with a NULL Pointer Dereferences vulnerabiity. If exploited would potentially cause disruption of the centralized monitoring of multiple beds and the visualization of alarms.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter