Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0141-2025

GE Healthcare Giraffe Incubator Carestation CS1, Infant Incubator. Model Numbers: 1) 2082844-002-01040417; 2) 2082844-002-01044956; 3) 2082844-002-01058424; 4) 2082844-002-01069017; 5) 2082844-002-01089666; 6) 2082844-002-01090474; 7) 2082844-002-01091269; 8) 2082844-002-01093489; 9) 2082844-002-01113262; 10) 2082844-002-01113711; 11) 2082844-002-01123085; 12) 2082844-002-01123918; 13) 2082844-002-01128966; 14) 2082844-002-01133042; 15) 2082844-002-01146447; 16) 2082844-002-100205; 17) 2082844-002-100749; 18) 2082844-002-100752; 19) 2082844-002-100995; 20) 2082844-002-101425; 21) 2082844-002-523548; 22) 2082844-002-529928; 23) 2082844-002-594800; 24) 2082844-002-643544; 25) 2082844-002-925841; 26) 2082844-002-965540; 27) 2082844-002-982916; 28) 2082844-002-983420; 29) 2082844-002-993905; 30) 2082844-002-995018; 31) 2082844-002-995799; 32) 2082844-002-998943; Radiant infant warmer

Publicado el 6 de noviembre de 2024 · Inicio del retiro: 6 de junio de 2023

Estado reportado al momento de publicación: Completed — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week. Affected customers were notified in June 2023 to run incubators for a week to potentially off gas any residual airborne formaldehyde. The general scientific consensus is that sensory irritation of the eyes is the most sensitive endpoint with respect to exposure to airborne formaldehyde and is the adverse effect clinically experienced at the lowest concentration.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
DATEX--OHMEDA, INC.
Distribuidor
worldwide
Lote
1) 2082844-002-01040417, UDI: (01)00840682116855(21)TAAC70003(11)230106, (01)00840682116855(21)TAAC70004(11)230109, (01)00840682116855(21)TAAC70011(11)230110, (01)00840682116855(21)TAAC70012(11)230110, (01)00840682116855(21)TAAC70013(11)230111, (01)00840682116855(21)TAAC70015(11)230112, (01)00840682116855(21)TAAC70016(11)230112, (01)00840682116855(21)TAAC70020(11)230112, (01)00840682116855(21)TAAC70022(11)230112, (01)00840682116855(21)TAAC70024(11)230112, (01)00840682116855(21)TAAC70147(11)230317, (01)00840682116855(21)TAAC70148(11)230317, (01)00840682116855(21)TAAC70149(11)230317, (01)00840682116855(21)TAAC70150(11)230317, (01)00840682116855(21)TAAC70152(11)230320, (01)00840682116855(21)TAAC70153(11)230320, (01)00840682116855(21)TAAC70155(11)230322, (01)00840682116855(21)TAAC70158(11)230322, (01)00840682116855(21)TAAC70159(11)230323, (01)00840682116855(21)TAAC70160(11)230323, (01)00840682116855(21)TAAC70161(11)230323, (01)00840682116855(21)TAAC70162(11)230323, (01)00840682116855(21)TAAC70163(11)230324, (01)00840682116855(21)TAAC70164(11)230324, (01)00840682116855(21)TAAC70165(11)230324, (01)00840682116855(21)TAAC70166(11)230324, (01)00840682116855(21)TAAC70167(11)230327, (01)00840682116855(21)TAAC70169(11)230327, (01)00840682116855(21)TAAC70170(11)230327, (01)00840682116855(21)TAAC70171(11)230328, (01)00840682116855(21)TAAC70177(11)230330, (01)00840682116855(21)TAAC70178(11)230330, (01)00840682116855(21)TAAC70179(11)230331, (01)00840682116855(21)TAAC70181(11)230331, (01)00840682116855(21)TAAC70182(11)230331, (01)00840682116855(21)TAAC70184(11)230331, (01)00840682116855(21)TAAC70192(11)230405, (01)00840682116855(21)TAAC70193(11)230405, (01)00840682116855(21)TAAC70196(11)230405, (01)00840682116855(21)TAAC70197(11)230405, (01)00840682116855(21)TAAC70200(11)230407, (01)00840682116855(21)TAAC70201(11)230407, (01)00840682116855(21)TAAC70202(11)230407, (01)00840682116855(21)TAAC70203(11)230407, (01)00840682116855(21)TAAC70204(11)230407, (01)00840682116855(21)TAAC70205(11)230407, (01)00840682116855(21)TAAC70208(11)230411, (01)00840682116855(21)TAAC70209(11)230411, (01)00840682116855(21)TAAC70210(11)230411, (01)00840682116855(21)TAAC70211(11)230414, (01)00840682116855(21)TAAC70212(11)230414, (01)00840682116855(21)TAAC70213(11)230414, (01)00840682116855(21)TAAC70214(11)230414, (01)00840682116855(21)TAAC70215(11)230417, (01)00840682116855(21)TAAC70216(11)230417, (01)00840682116855(21)TAAC70217(11)230417, (01)00840682116855(21)TAAC70218(11)230417, (01)00840682116855(21)TAAC70223(11)230418, (01)00840682116855(21)TAAC70224(11)230418, (01)00840682116855(21)TAAC70225(11)230418, (01)00840682116855(21)TAAC70226(11)230418, (01)00840682116855(21)TAAC70227(11)230418, (01)00840682116855(21)TAAC70228(11)230418, (01)00840682116855(21)TAAC70229(11)230419, (01)00840682116855(21)TAAC70242(11)230421, (01)00840682116855(21)TAAC70243(11)230421, (01)00840682116855(21)TAAC70244(11)230421, (01)00840682116855(21)TAAC70245(11)230421, (01)00840682116855(21)TAAC70246(11)230421, (01)00840682116855(21)TAAC70247(11)230421; 2) 2082844-002-01044956, UDI: (01)008406821168552(21)TAA23140007SA(11)230404, (01)008406821168552(21)TAA23140008SA(11)230404; 3) 2082844-002-01058424, UDI: (01)00840682116855(21)TAAC70142(11)230213; 4) 2082844-002-01069017, UDI: (01)008406821168552(21)TAA23110001SA(11)230316, (01)008406821168552(21)TAA23110002SA(11)230316, (01)008406821168552(21)TAA23110003SA(11)230316; 5) 2082844-002-01089666, UDI: (01)00840682116855(21)TAAC70124(11)230201; 6) 2082844-002-01090474, UDI: (01)00840682116855(21)TAAC70126(11)230201, (01)00840682116855(21)TAAC70128(11)230206, (01)00840682116862(21)TAAC70132(11)230206, (01)00840682116855(21)TAAC70133(11)230206, (01)00840682116855(21)TAAC70125(11)230201, (01)00840682116855(21)TAAC70127(11)230206, (01)00840682116855(21)TAAC70129(11)230206, (01)00840682116855(21)TAAC70131(11)230206, (01)00840682116855(21)TAAC70136(11)230206, (01)00840682116855(21)TAAC70137(11)230206; 7) 2082844-002-01091269, UDI: (01)0
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States