Automated Impella Controller, Packaged. Primary user control interface for the Impella Catheter, catheter-based heart pump.

Publicado el 4 de noviembre de 2026 · Inicio del retiro: 18 de septiembre de 2026

Resumen

FDA registra el retiro Z-0208-2027 de dispositivos médicos de Clase I, publicado el 4 de noviembre de 2026 e iniciado el 18 de septiembre de 2026. Empresa responsable: Abiomed, Inc.; Estado: En curso; Cantidad reportada: 428 units; Lote: Product Code: 0042-0000-US, 0042-0000-UK, 0042-0000-EU, 0042-0000-IN, 0042-0000-CA; UDI:00813502010022, 00813502011296, 00813502011289, 00813502011418, 00813502011272; Serial No. IC1014, IC1029, IC1038, IC1043, IC1042, IC1061, IC1075, IC1096, IC1122, IC1190, IC1202, IC1203, IC1224, IC1231, IC1238, IC1269, IC1289, IC1446, IC1612, IC1741, IC1771, IC1852, IC1853, IC1708, IC1964, IC2071, IC2069, IC2154, IC2227, IC2314, IC2315, IC2029, IC2321, IC2428, IC2429, IC2521, IC2561, IC2562, IC2575, IC2664, IC2665, IC2691, IC2692, IC2739, IC2782, IC2840, IC2841, IC2848, IC2891, IC2976, IC2977, IC3009, IC3010, IC3154, IC3155, IC3244, IC3246, IC3342, IC3398, IC3347, IC3351, IC3512, IC3609, IC3714, IC3715, IC3772, IC3773, IC3806, IC3822, IC3823, IC3919, IC3975, IC3976, IC4041, IC4042, IC4045, IC4060, IC4250, IC4259, IC4295, IC4339, IC4616, IC4640, IC4641, IC4708, IC4831, IC4832, IC4884, IC5120, IC5151, IC5201, IC5056, IC5549, IC5605, IC5661, IC5662, IC5598, IC5720, IC5721, IC5745, IC5749, IC5750, IC5828, IC5984, IC6042, IC6321, IC6322, IC6328, IC6329, IC6359, IC6360, IC6403, IC6596, IC6597, IC6622, IC6623, IC6632, IC6637, IC6673, IC6825, IC6846, IC6834, IC6835, IC1387, IC1869, IC1909, IC2277, IC2278, IC2871, IC2872, IC3293, IC3294, IC3295, IC3674, IC3675, IC3784, IC3785, IC4741, IC5219, IC5220, IC5553, IC5911, IC5912, IC5926, IC5927, IC6289, IC6286, IC6287, IC6540, IC6583, IC6584, IC6585, IC6845, IC6925, IC7057, IC7074, IC7073, IC7441, IC7442, IC7591, IC1285, IC1416, IC3719, IC3720, IC3776, IC1825, IC1876, IC2175, IC2857, IC2996, IC3296, IC3486, IC3488, IC3890, IC4055, IC4323, IC4681, IC4682, IC4892, IC4984, IC4987, IC4973, IC5048, IC5421, IC5442, IC5464, IC5494, IC5687, IC5611, IC5722, IC5818, IC6065, IC6536, IC6582, IC6588, IC6589, IC6688, IC6694, IC6706, IC6707, IC6728, IC6738, IC6739, IC6740, IC6877, IC7388, IC7389, IC7390, IC7647, IC7649, IC7651, IC7652, IC7776, IC7777, IC5833, IC7444, IC7445, IC7446, IC7447, IC1284, IC1545, IC2368, IC2989, IC4331, IC4438, IC4856, IC2099, IC2823, IC3668, IC4099, IC4218, IC4611, IC4932, IC5490, IC5493, IC7407, IC7675, IC7686, IC7691, IC7692, IC7721, IC7809, IC3575, IC3576, IC3837, IC4065, IC4066, IC4451, IC4452, IC6330, IC3131, IC3132, IC2234, IC2469, IC3331, IC3669, IC3777, IC3828, IC4150, IC1396, IC1397, IC1724, IC1725, IC3050, IC5755, IC5756, IC3301, IC3302, IC3312, IC4675, IC4676, IC1577, IC1594, IC2877, IC6708, IC6806, IC5184, IC5218, IC7443, IC7585, IC5995, IC5996, IC5997, IC5998, IC5999, IC6001, IC6252, IC6253, IC6254, IC6139, IC6476, IC6477, IC6478, IC6475, IC7079, IC7080, IC7408, IC7409, IC7410, IC7411, IC3174, IC3166, IC3167, IC3213, IC3223, IC3475, IC4441, IC4525, IC5215, IC5910, IC7087, IC1280, IC1283, IC1345, IC1411, IC1398, IC2240, IC2557, IC3017, IC3178, IC3179, IC3322, IC3319, IC3490, IC3660, IC3813, IC4232, IC4237, IC4240, IC4242, IC4469, IC4626, IC4737, IC1586, IC1677, IC2021, IC2206, IC2245, IC3016, IC3049, IC3494, IC4329, IC4358, IC4563, IC4723, IC5757, IC6079, IC6080, IC6431, IC6580, IC6581, IC7719, IC3704, IC3705, IC5055, IC1766, IC5435, IC2451, IC2452, IC2453, IC2459, IC2460, IC2461, IC2449, IC1001, IC1179, IC1346, IC1481, IC1595, IC1715, IC1769, IC1790, IC1791, IC1911, IC1913, IC1912, IC2033, IC2035, IC2687, IC3661, IC3662, IC3802, IC4944, IC5034, IC5821, IC5824, IC6082, IC6083, IC6633, IC6636, IC7695, IC7716, IC9413, IC9414, IC9416, IC2442, IC2444, IC2443, IC2476, IC2477, IC2478, IC2479, IC2471, IC3127, IC3128, IC3129, IC3130, IC3306, IC3307, IC3308, IC3309, IC3656, IC3658, IC3663, IC3664, IC3665, IC3666, IC3783, IC3812, IC3888, IC3887, IC4068, IC4069, IC4257, IC4507, IC7578, IC3699, IC3700. La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. 2. Potential for the rotary push button (RPB) sticking and unreliable actuation impacting operator ability to select and confirm menu entries. This can result in loss of control of the console. 3. Potential AIC instability that could result in AIC start-up failure, inability to drive the pump, and the need for console exchange. 4. Software defect related to calculator process crash, leading to Negative Central Venous Pressure (CVP) values present in Pulmonary Artery Pressure (PAP) calculations.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Abiomed, Inc.
Distribuidor
Worldwide Distribution: US (nationwide) and OUS (foreign) to countries: Belgium, Brazil, Canada, Colombia, Croatia, Czech Republic, Finland, France, Germany, Hong Kong, India, Ireland, Italy, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Pakistan, Panama, Poland, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom
Lote
Product Code: 0042-0000-US, 0042-0000-UK, 0042-0000-EU, 0042-0000-IN, 0042-0000-CA; UDI:00813502010022, 00813502011296, 00813502011289, 00813502011418, 00813502011272; Serial No. IC1014, IC1029, IC1038, IC1043, IC1042, IC1061, IC1075, IC1096, IC1122, IC1190, IC1202, IC1203, IC1224, IC1231, IC1238, IC1269, IC1289, IC1446, IC1612, IC1741, IC1771, IC1852, IC1853, IC1708, IC1964, IC2071, IC2069, IC2154, IC2227, IC2314, IC2315, IC2029, IC2321, IC2428, IC2429, IC2521, IC2561, IC2562, IC2575, IC2664, IC2665, IC2691, IC2692, IC2739, IC2782, IC2840, IC2841, IC2848, IC2891, IC2976, IC2977, IC3009, IC3010, IC3154, IC3155, IC3244, IC3246, IC3342, IC3398, IC3347, IC3351, IC3512, IC3609, IC3714, IC3715, IC3772, IC3773, IC3806, IC3822, IC3823, IC3919, IC3975, IC3976, IC4041, IC4042, IC4045, IC4060, IC4250, IC4259, IC4295, IC4339, IC4616, IC4640, IC4641, IC4708, IC4831, IC4832, IC4884, IC5120, IC5151, IC5201, IC5056, IC5549, IC5605, IC5661, IC5662, IC5598, IC5720, IC5721, IC5745, IC5749, IC5750, IC5828, IC5984, IC6042, IC6321, IC6322, IC6328, IC6329, IC6359, IC6360, IC6403, IC6596, IC6597, IC6622, IC6623, IC6632, IC6637, IC6673, IC6825, IC6846, IC6834, IC6835, IC1387, IC1869, IC1909, IC2277, IC2278, IC2871, IC2872, IC3293, IC3294, IC3295, IC3674, IC3675, IC3784, IC3785, IC4741, IC5219, IC5220, IC5553, IC5911, IC5912, IC5926, IC5927, IC6289, IC6286, IC6287, IC6540, IC6583, IC6584, IC6585, IC6845, IC6925, IC7057, IC7074, IC7073, IC7441, IC7442, IC7591, IC1285, IC1416, IC3719, IC3720, IC3776, IC1825, IC1876, IC2175, IC2857, IC2996, IC3296, IC3486, IC3488, IC3890, IC4055, IC4323, IC4681, IC4682, IC4892, IC4984, IC4987, IC4973, IC5048, IC5421, IC5442, IC5464, IC5494, IC5687, IC5611, IC5722, IC5818, IC6065, IC6536, IC6582, IC6588, IC6589, IC6688, IC6694, IC6706, IC6707, IC6728, IC6738, IC6739, IC6740, IC6877, IC7388, IC7389, IC7390, IC7647, IC7649, IC7651, IC7652, IC7776, IC7777, IC5833, IC7444, IC7445, IC7446, IC7447, IC1284, IC1545, IC2368, IC2989, IC4331, IC4438, IC4856, IC2099, IC2823, IC3668, IC4099, IC4218, IC4611, IC4932, IC5490, IC5493, IC7407, IC7675, IC7686, IC7691, IC7692, IC7721, IC7809, IC3575, IC3576, IC3837, IC4065, IC4066, IC4451, IC4452, IC6330, IC3131, IC3132, IC2234, IC2469, IC3331, IC3669, IC3777, IC3828, IC4150, IC1396, IC1397, IC1724, IC1725, IC3050, IC5755, IC5756, IC3301, IC3302, IC3312, IC4675, IC4676, IC1577, IC1594, IC2877, IC6708, IC6806, IC5184, IC5218, IC7443, IC7585, IC5995, IC5996, IC5997, IC5998, IC5999, IC6001, IC6252, IC6253, IC6254, IC6139, IC6476, IC6477, IC6478, IC6475, IC7079, IC7080, IC7408, IC7409, IC7410, IC7411, IC3174, IC3166, IC3167, IC3213, IC3223, IC3475, IC4441, IC4525, IC5215, IC5910, IC7087, IC1280, IC1283, IC1345, IC1411, IC1398, IC2240, IC2557, IC3017, IC3178, IC3179, IC3322, IC3319, IC3490, IC3660, IC3813, IC4232, IC4237, IC4240, IC4242, IC4469, IC4626, IC4737, IC1586, IC1677, IC2021, IC2206, IC2245, IC3016, IC3049, IC3494, IC4329, IC4358, IC4563, IC4723, IC5757, IC6079, IC6080, IC6431, IC6580, IC6581, IC7719, IC3704, IC3705, IC5055, IC1766, IC5435, IC2451, IC2452, IC2453, IC2459, IC2460, IC2461, IC2449, IC1001, IC1179, IC1346, IC1481, IC1595, IC1715, IC1769, IC1790, IC1791, IC1911, IC1913, IC1912, IC2033, IC2035, IC2687, IC3661, IC3662, IC3802, IC4944, IC5034, IC5821, IC5824, IC6082, IC6083, IC6633, IC6636, IC7695, IC7716, IC9413, IC9414, IC9416, IC2442, IC2444, IC2443, IC2476, IC2477, IC2478, IC2479, IC2471, IC3127, IC3128, IC3129, IC3130, IC3306, IC3307, IC3308, IC3309, IC3656, IC3658, IC3663, IC3664, IC3665, IC3666, IC3783, IC3812, IC3888, IC3887, IC4068, IC4069, IC4257, IC4507, IC7578, IC3699, IC3700.
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Automated Impella Controller, Packaged. Primary user control interface for the Impella Catheter, catheter-based heart pump. fue retirado del mercado?

Sí. FDA publicó el retiro Z-0208-2027 el 4 de noviembre de 2026. Peligro reportado: Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. 2. Potential for the rotary push button (RPB) sticking and unreliable actuation impacting operator ability to select and confirm menu entries. This can result in loss of control of the console. 3. Potential AIC instability that could result in AIC start-up failure, inability to drive the pump, and the need for console exchange. 4. Software defect related to calculator process crash, leading to Negative Central Venous Pressure (CVP) values present in Pulmonary Artery Pressure (PAP) calculations.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter