t:slim X2 insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014070, 1014071, 1014072, 1014073, and User Guide when using Dexcom G7 sensor; Tandem Mobi insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014801, 1014811, 1014813, and User Guide when using Dexcom G7 sensor
Publicado el 29 de octubre de 2025 · Inicio del retiro: 5 de agosto de 2025
Resumen
FDA registra el retiro Z-0232-2026 de dispositivos médicos de Clase II, publicado el 29 de octubre de 2025 e iniciado el 5 de agosto de 2025. Empresa responsable: Tandem Diabetes Care, Inc.; Estado: En curso; Cantidad reportada: 122,958; Lote: t:slim X2, UDI-DI: 00389152407012, 00389152407111, 00389152407210, 00389152407319 with software versions 7.9.0.1 and 7.10.1, User Guide versions not including and prior to: AW-1014562 Revision D; Tandem Mobi, UDI-DI: 00389152480114, 00389152481111, 00389152481319 with software version 7.9.0.1, User Guide versions not including and prior to: AW-1014938 Revision C. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an unexpected automatic insulin correction bolus (autobolus ) which could result in hypoglycemia.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Tandem Diabetes Care, Inc.
- Distribuidor
- US nationwide distribution including Puerto Rico.
- Lote
- t:slim X2, UDI-DI: 00389152407012, 00389152407111, 00389152407210, 00389152407319 with software versions 7.9.0.1 and 7.10.1, User Guide versions not including and prior to: AW-1014562 Revision D; Tandem Mobi, UDI-DI: 00389152480114, 00389152481111, 00389152481319 with software version 7.9.0.1, User Guide versions not including and prior to: AW-1014938 Revision C
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDATandem Mobi Insulin Pump, Models: 1004004, 1010750, 1012676, 1012719, 1013655, 1013656, 1013501, 1013700,11 de noviembre de 2026
- FDABrand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog7 de octubre de 2026
- FDAFuse ULTRA Foot Plating System Instructions for Use. Models: OS200400, OS200401, OS328017, OS200020L,29 de octubre de 2025
- FDAOral Probe (Product code 02893-000), accessory used with Welch Allyn SureTemp Plus 690 (Product codes29 de octubre de 2025
- FDAShanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System29 de octubre de 2025
- FDAt:connect mobile app used in conjunction with t:slim X2 insulin pump with Control-IQ technology15 de mayo de 2024
Preguntas sobre este retiro
¿t:slim X2 insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014070, 1014071, 1014072, 1014073, and User Guide when using Dexcom G7 sensor; Tandem Mobi insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014801, 1014811, 1014813, and User Guide when using Dexcom G7 sensor fue retirado del mercado?
Sí. FDA publicó el retiro Z-0232-2026 el 29 de octubre de 2025. Peligro reportado: A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an unexpected automatic insulin correction bolus (autobolus ) which could result in hypoglycemia.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter