Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0252-2027

Direxion Torqueable Microcatheter, NGMC/FATHOM/021/BERN/1RO/155; Material Number: M001195650

Publicado el 28 de octubre de 2026 · Inicio del retiro: 16 de septiembre de 2026

Resumen

FDA registra el retiro Z-0252-2027 de dispositivos médicos de Clase II, publicado el 28 de octubre de 2026 e iniciado el 16 de septiembre de 2026. Empresa responsable: Boston Scientific Cork, Ltd.; Estado: En curso; Cantidad reportada: 860; Lote: Material Number: M001195650; UDI-DI: 08714729839644; Batch Numbers: 37886794, 37957747, 38022670, 38049176, 38149828, 38175798, 38224585, 38259485, 38323409, 38506199, 38531120, 38560457, 38595646, 38595647, 38745191, 38769968, 38769971, 38916516, 38970729, 38990859, 38990882, 39027327, 39141260, 39189845, 40466132, 40466133, 40466134, 40485781, 40493314, 40502331, 40502332, 40502333, 40504594, 40507434, 40507435, 40531377, 40535710, 40535711, 40537146. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Boston Scientific Cork, Ltd.
Distribuidor
Worldwide distribution US Nationwide and territory of Puerto Rico. The countries of Albania, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Mexico, Moldova, Netherlands, New Zealand, Norway, Panama, PERU, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, T®rkiye, UAE, United Kingdom.
Lote
Material Number: M001195650; UDI-DI: 08714729839644; Batch Numbers: 37886794, 37957747, 38022670, 38049176, 38149828, 38175798, 38224585, 38259485, 38323409, 38506199, 38531120, 38560457, 38595646, 38595647, 38745191, 38769968, 38769971, 38916516, 38970729, 38990859, 38990882, 39027327, 39141260, 39189845, 40466132, 40466133, 40466134, 40485781, 40493314, 40502331, 40502332, 40502333, 40504594, 40507434, 40507435, 40531377, 40535710, 40535711, 40537146;
Clasificación de riesgo
Class II
Área (FDA)
device
País
Ireland

Retiros relacionados

Preguntas sobre este retiro

¿Direxion Torqueable Microcatheter, NGMC/FATHOM/021/BERN/1RO/155; Material Number: M001195650 fue retirado del mercado?

Sí. FDA publicó el retiro Z-0252-2027 el 28 de octubre de 2026. Peligro reportado: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter