Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0263-2014

THUNDER¿ Extra Support Guidewire; Models: THNDR190J, THNDR190S, THNDR300J, THNDR300S Medtronic guide wires are steerable guide wires that are used for the introduction and placement of diagnostics or interventional devices in the coronary and peripheral vasculature, and may be used to reach and cross a target lesion. Medtronic guide wires are not intended for cerebral vasculature. Medtronic steerable exchange guide wires are used to facilitate the substitution of one diagnostic or interventional device for another.

Publicado el 27 de noviembre de 2013 · Inicio del retiro: 21 de octubre de 2013

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Potential for PTFE (polytetrafluroethylene) coating to delaminate and detach from Guidewire.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Medtronic Vascular
Distribuidor
Worldwide distribution: US (Nationwide) including states of: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV; and countries of: Albania, Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Georgia, Germany, Greece, Guyana, Honduras, Hong Kong,Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Libya, Luxembourg, Malaysia, Mexico, Montenegro, Nepal, Netherlands, New Zealand, Nepal, Nicaragua, Norway, Oman, Poland, Portugal Puerto Rico, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, United Kingdom, Uruguay, and Yemen.
Lote
G13A00117 G13A00735 G13A00792 G13A00794 G13A01324 GV259 GV261 GV276 GV287 GV847 GV848 GV849 GV851 GV852 GW690 GW758 GU141 GU959
Clasificación de riesgo
Class I
Área (FDA)
device