Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0323-2017

Visualase Body Accessory Kit PN 020-3000. The individual components within the Visualase Body Accessory Kit are packaged in Tyvek pouches. The packaged devices are then placed into a cardboard box which is labeled with the Body Accessory Kit label. Labeled sterile. The visualization stylet is constructed of titanium and used during Visualase laser ablation procedures by inserting it into intracorporeal catheter lumens for improved catheter stiffness and MRI visualization during procedures.

Publicado el 9 de noviembre de 2016 · Inicio del retiro: 1 de septiembre de 2016

Resumen

FDA registra el retiro Z-0323-2017 de dispositivos médicos de Clase II, publicado el 9 de noviembre de 2016 e iniciado el 1 de septiembre de 2016. Empresa responsable: Medtronic Navigation, Inc.; Estado: Finalizado; Cantidad reportada: 100 total units; Lote: Lot number 151591. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Medtronic Navigation, Inc. announces a voluntary field action for the Visualase Visualization Stylets and Visualase Body Accessory Kit because when the Visualization Stylet is placed in the magnetic field of a magnetic resonance imaging system it pulls and/or bends toward the field.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Medtronic Navigation, Inc.
Distribuidor
Worldwide Distribution - US including CA, GA, FL, and TX. Internationally to Australia.
Lote
Lot number 151591
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Visualase Body Accessory Kit PN 020-3000. The individual components within the Visualase Body Accessory Kit are packaged in Tyvek pouches. The packaged devices are then placed into a cardboard box which is labeled with the Body Accessory Kit label. Labeled sterile. The visualization stylet is constructed of titanium and used during Visualase laser ablation procedures by inserting it into intracorporeal catheter lumens for improved catheter stiffness and MRI visualization during procedures. fue retirado del mercado?

Sí. FDA publicó el retiro Z-0323-2017 el 9 de noviembre de 2016. Peligro reportado: Medtronic Navigation, Inc. announces a voluntary field action for the Visualase Visualization Stylets and Visualase Body Accessory Kit because when the Visualization Stylet is placed in the magnetic field of a magnetic resonance imaging system it pulls and/or bends toward the field.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter