Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0367-2025

Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Non-coated, 70mm - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. Catalog Number: 0703-047-001

Publicado el 20 de noviembre de 2024 · Inicio del retiro: 4 de octubre de 2024

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Stryker Corporation
Distribuidor
Worldwide distribution - US Nationwide and the countries of Canada, Japan, Australia.
Lote
GTIN: 07613327117219 Lot Number: 64117291
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States