Stryker Neptune SafeAir Smoke Evacuation Pencil Coated Push-Button, Bulk Non-sterile - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. The pencil enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit to the operative site for the desired surgical effect. Catalog Number: 0703-046-002
Publicado el 20 de noviembre de 2024 · Inicio del retiro: 4 de octubre de 2024
Resumen
FDA registra el retiro Z-0370-2025 de dispositivos médicos de Clase II, publicado el 20 de noviembre de 2024 e iniciado el 4 de octubre de 2024. Empresa responsable: Stryker Corporation; Estado: En curso; Cantidad reportada: 403,730 eaches; Lote: GTIN: 07613327379488 Lot Numbers: 53864143 57458172 58761873 63837701 64258041 65777003 66126554 66751061 68311144 54406823 57458173 58761874 63837702 64258042 65777004 66487383 67200500 68311145 54406824 57458174 60308898 63866394 64258044 65954742 66487384 67200501 68672590 54406825 57458175 60308899 63866395 64850548 65954743 66487385 67200502 68728239 54406826 57694644 62291709 64052771 64850549 65954744 66487386 67715714 68728240 54406827 57733684 62377382 64112622 64968511 65954745 66751059 67787217 54406828 58009712 62537206 64112623 64968512 66126553 66751060 68200024. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Stryker Corporation
- Distribuidor
- Worldwide distribution - US Nationwide and the countries of Canada, Japan, Australia.
- Lote
- GTIN: 07613327379488 Lot Numbers: 53864143 57458172 58761873 63837701 64258041 65777003 66126554 66751061 68311144 54406823 57458173 58761874 63837702 64258042 65777004 66487383 67200500 68311145 54406824 57458174 60308898 63866394 64258044 65954742 66487384 67200501 68672590 54406825 57458175 60308899 63866395 64850548 65954743 66487385 67200502 68728239 54406826 57694644 62291709 64052771 64850549 65954744 66487386 67715714 68728240 54406827 57733684 62377382 64112622 64968511 65954745 66751059 67787217 54406828 58009712 62537206 64112623 64968512 66126553 66751060 68200024
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDASTRYKER PKG., STRYKEFLOW SUCTION/IRRIGATOR PS, Catalog Number 025007054012 de agosto de 2026
- FDASTRYKER AHTO DUAL SPIKE TUBE SET PACKAGING, Catalog Number 025007064012 de agosto de 2026
- FDASTRYKER AHTO TUBE SET WITH TIP PACKAGING, Catalog Number 025007062012 de agosto de 2026
- FDASTRYKER AHTO TUBE SET PACKAGING, Catalog Number 025007060012 de agosto de 2026
- FDAStryker Neptune SafeAir Smoke Evacuation Pencil Non coated 70 mm Blade Push-Button - Designed for general20 de noviembre de 2024
- FDAThe EVSRF catheter is a sterile, single-use disposable medical device for endovenous radiofrequency ablation,20 de noviembre de 2024
Preguntas sobre este retiro
¿Stryker Neptune SafeAir Smoke Evacuation Pencil Coated Push-Button, Bulk Non-sterile - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. The pencil enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit to the operative site for the desired surgical effect. Catalog Number: 0703-046-002 fue retirado del mercado?
Sí. FDA publicó el retiro Z-0370-2025 el 20 de noviembre de 2024. Peligro reportado: There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter