FDAFDA · Alimentos y medicamentos·N.º oficial Z-0391-2023
Carefusion V. Mueller" CUSHING BAYONET BIPOLAR FORCEPS, 2.0MM TIP, INSULATED OVERALL LENGTH 7-1/2" (190MM), REF F-2004
Publicado el 14 de diciembre de 2022 · Inicio del retiro: 2 de noviembre de 2022
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
BD has determined that the Instructions For Use (IFU) provided by BD is missing certain content that is contained in the manufacturer's IFU. This missing content is related to the interface with the device power supply, as well as cleaning and maintenance instructions. Refer to the attached V. Mueller" Bipolar Forcep IFU.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Carefusion 2200 Inc
- Distribuidor
- Nationwide, Austria, Canada, Egypt, Germany, Hong Kong, Israel, Philippines, Saudi Arabia, South Africa, South Korea, Thailand
- Lote
- UDI/DI 10885403040917, ALL LOTS
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
¿Vendes productos como este?
Recalert vigila todo tu catálogo contra retiros como este, automáticamente.