OmniGuide BP-ROBOTIC ( BeamPath Robotic), Single Use Cat. # BP-ROBOTIC The OmniGuide Fiber is indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues including intra-oral tissues. It is indicated in the medical specialties of general and plastic surgery, oral/maxillofacial surgery, dentistry, dermatology, gynecology, otorhinolaryngology, gastroenterology, neurosurgery, urology, and pulmonology, and can be used in open surgical procedures as well as endoscopic minimally invasive procedures in conjunction with rigid or flexible endoscopes, such as in laryngoscopy, gastroscopy, colonoscopy, laparoscopy, thoracoscopy, hysteroscopy and bronchoscopy
Publicado el 12 de diciembre de 2012 · Inicio del retiro: 5 de noviembre de 2012
Resumen
FDA registra el retiro Z-0453-2013 de dispositivos médicos de Clase II, publicado el 12 de diciembre de 2012 e iniciado el 5 de noviembre de 2012. Empresa responsable: OmniGuide, Inc.; Estado: Finalizado; Cantidad reportada: 45 units; Lote: Lot Numbers: LA121003AO-P1 LA121003AQ-P1 LA121001CB-P1 LA121003AL-P1 LA121002AC-P1 LA121001CC-P1 LA121001BZ-P1 LA121001BZ-P1 LA121001AR-P1 LA121001AQ-P1 LA121001CA-P1 LA120926BA-P1 LA120926AS-P1 LA120919BG-P1 LA120920AF-P1 LA120924AG-P1 :. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Failure of glue joint, allowed a portion of the single use device to shift inside, may separate and fall into the patient
Qué hacer
Voluntary: Firm initiated · E-Mail
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- OmniGuide, Inc.
- Distribuidor
- Nationwide Distribution including AZ, FL, MA, NY, OH, and OR.
- Lote
- Lot Numbers: LA121003AO-P1 LA121003AQ-P1 LA121001CB-P1 LA121003AL-P1 LA121002AC-P1 LA121001CC-P1 LA121001BZ-P1 LA121001BZ-P1 LA121001AR-P1 LA121001AQ-P1 LA121001CA-P1 LA120926BA-P1 LA120926AS-P1 LA120919BG-P1 LA120920AF-P1 LA120924AG-P1 :
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDADimension Vista(R) IRON Flex(R) reagent cartridge (K3085) The IRON method is an in vitro diagnostic test for12 de diciembre de 2012
- FDANeptune® 2 Rover - ULTRA ~120V ~ 60 Hz 12A REF 0702-001-000, Rx, Caution, consult accompanying documents12 de diciembre de 2012
- FDAThe product has the Catalogue/Part number 60010. Olympic Cool-Cap System is a selective head cooling system12 de diciembre de 2012
- FDANeptune® 2 Rover - ULTRA ~230V ~ 50 Hz 12A REF 0702-002-000, Rx Caution, consult accompanying documents12 de diciembre de 2012
- FDATomales Bay Foods brand Artisan Cheese in PARTIAL WHEELS of Cheese, from cheese makers in US (VA, VT, Ca, WI,12 de diciembre de 2012
- FDA24 Fr 8mm Sft Flow Str Wire W/L The Sarns Aortic Cannula is indicated for use in perfusion of the ascending12 de diciembre de 2012
Preguntas sobre este retiro
¿OmniGuide BP-ROBOTIC ( BeamPath Robotic), Single Use Cat. # BP-ROBOTIC The OmniGuide Fiber is indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues including intra-oral tissues. It is indicated in the medical specialties of general and plastic surgery, oral/maxillofacial surgery, dentistry, dermatology, gynecology, otorhinolaryngology, gastroenterology, neurosurgery, urology, and pulmonology, and can be used in open surgical procedures as well as endoscopic minimally invasive procedures in conjunction with rigid or flexible endoscopes, such as in laryngoscopy, gastroscopy, colonoscopy, laparoscopy, thoracoscopy, hysteroscopy and bronchoscopy fue retirado del mercado?