Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0465-2022

Disposable Subdermal Needle Electrode, Corkscrew, Single Use, Sterile; Model Numbers TE/S50715-001-C, TE/S50725-001C, and TE/S50715-001.

Publicado el 19 de enero de 2022 · Inicio del retiro: 8 de diciembre de 2021

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

There is a high risk of the blue hub detaching from the corkscrew needle, resulting the the needle remaining in the patient's scalp.

Qué hacer

Voluntary: Firm initiated · E-Mail

Detalles del retiro

Fabricante
Technomed Europe
Distribuidor
US Nationwide distribution in the states of California, Florida, and Ohio.
Lote
Read as "Lot Number (UDI-DI, Product Number)": 048565 (08718375861707, TE/S50715-004), 048670 (08718375867549, TE/S50715-001-C), 048949 (08718375867549, TE/S50715-001-C), 049013 (08718375865439, SO-ELCS20), 049013 (08718375865439, SO-ELCS20), 049014 (08718375867549, TE/S50715-001-C), 049014 (08718375867549, TE/S50715-001-C), 049165 (08718375865439, SO-ELCS20), 049166 (08718375867853, TE/S50725-001-C), 049166 (08718375867853, TE/S50725-001-C), 049166 (08718375867853, TE/S50725-001-C), 049166 (08718375867853, TE/S50725-001-C), 049167 (08718375867549, TE/S50715-001-C), 049167 (08718375867549, TE/S50715-001-C), 049167 (08718375867549, TE/S50715-001-C), 049167 (08718375867549, TE/S50715-001-C), 049188 (08718375861707, TE/S50715-004), 049188 (08718375861707, TE/S50715-004), 049188 (08718375861707, TE/S50715-004), 049188 (08718375861707, TE/S50715-004), 049188 (08718375861707, TE/S50715-004), 049188 (08718375861707, TE/S50715-004), 049189 (08718375861707, TE/S50715-004), 049189 (08718375861707, TE/S50715-004), 049189 (08718375861707, TE/S50715-004), 049189 (08718375861707, TE/S50715-004), 049189 (08718375861707, TE/S50715-004), 049190 (08718375861691, TE/S50715-003), 049190 (08718375861691, TE/S50715-003), 049190 (08718375861691, TE/S50715-003), 049190 (08718375861691, TE/S50715-003), 049190 (08718375861691, TE/S50715-003), 049191 (08718375861684, TE/S50715-002), 049191 (08718375861684, TE/S50715-002), 049191 (08718375861684, TE/S50715-002), 049191 (08718375861684, TE/S50715-002), 049192 (08718375861547, TE/S50715-001), 049192 (08718375861547, TE/S50715-001), 049192 (08718375861547, TE/S50715-001), 049192 (08718375861547, TE/S50715-001), 049193 (08718375866863, TE/S50725-002), 049193 (08718375866863, TE/S50725-002), 049193 (08718375866863, TE/S50725-002), 049193 (08718375866863, TE/S50725-002), 049328 (08718375861707, TE/S50715-004), 049328 (08718375861707, TE/S50715-004), 049328 (08718375861707, TE/S50715-004), 049343 (08718375861691, TE/S50715-003), 049343 (08718375861691, TE/S50715-003), 049343 (08718375861691, TE/S50715-003), 049343 (08718375861691, TE/S50715-003), 049343 (08718375861691, TE/S50715-003), 049429 (08718375861547, TE/S50715-001), 049429 (08718375861547, TE/S50715-001), 049429 (08718375861547, TE/S50715-001), 049429 (08718375861547, TE/S50715-001), 049429 (08718375861547, TE/S50715-001), 049429 (08718375861547, TE/S50715-001), 049429 (08718375861547, TE/S50715-001), 049429 (08718375861547, TE/S50715-001), 049429 (08718375861547, TE/S50715-001), 049429 (08718375861547, TE/S50715-001), 049429 (08718375861547, TE/S50715-001), 049429 (08718375861547, TE/S50715-001), 049429 (08718375861547, TE/S50715-001), 049443 (08718375867549, TE/S50715-001-C), 049443 (08718375867549, TE/S50715-001-C), 049502 (08718375867549, TE/S50715-001-C), 049528 (08718375861547, TE/S50715-001), 049528 (08718375861547, TE/S50715-001), 049528 (08718375861547, TE/S50715-001), 049528 (08718375861547, TE/S50715-001), 049528 (08718375861547, TE/S50715-001), 049549 (08718375867549, TE/S50715-001-C), 049552 (08718375867549, TE/S50715-001-C), 049552 (08718375867549, TE/S50715-001-C), 049603 (08718375861691, TE/S50715-003), 049662 (08718375861547, TE/S50715-001), 049662 (08718375861547, TE/S50715-001), 049662 (08718375861547, TE/S50715-001), 049662 (08718375861547, TE/S50715-001), 049662 (08718375861547, TE/S50715-001), 049710 (08718375867549, TE/S50715-001-C), 049710 (08718375867549, TE/S50715-001-C), 049735 (08718375861707, TE/S50715-004), 049735 (08718375861707, TE/S50715-004), 049736 (08718375867549, TE/S50715-001-C), 049802 (08718375867549, TE/S50715-001-C), 049847 (08718375865439, SO-ELCS20), 049847 (08718375865439, SO-ELCS20), 049855 (08718375861691, TE/S50715-003), 049855 (08718375861691, TE/S50715-003), 049855 (08718375861691, TE/S50715-003), 049858 (08718375861707, TE/S50715-004), 049858 (08718375861707, TE/S50715-004), 049861 (08718375867549, TE/S50715-001-C), 049871 (08718375866863, TE/S50725-002), 049878 (08718375867549, TE/S50715-001-C), 04992
Clasificación de riesgo
Class II
Área (FDA)
device
País
Netherlands

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta