NeuViz 16 Multi-Sliced CT Scanner System, part number 989605858501. NeuViz Dual series CT Scanner System, part number 989605651321 The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously.
Publicado el 12 de diciembre de 2012 · Inicio del retiro: 18 de octubre de 2012
Resumen
FDA registra el retiro Z-0466-2013 de dispositivos médicos de Clase II, publicado el 12 de diciembre de 2012 e iniciado el 18 de octubre de 2012. Empresa responsable: Philips And Neusoft Medical Systems Co., Ltd.; Estado: Finalizado; Cantidad reportada: 20 units of NeuViz 16 CT Scanner Systems and 25 units of NeuViz Dual series; Lote: NeuViz 16 system Serial numbers: N16E090002, N16E090003, N16E090004, N16E090007, N16E090008, N16E090010, N16E100015, N16E100017, N16E100023, N16E110016, N16E110023, N16E110026, N16E110027, N16E110040, N16E110053, N16E120016, N16E120022, N16E120020, N16E120023, N16E120024. Neuviz Dual Serial numbers: 400562, 400563, 400626, NDH009EI, NDH011EI, NDH012EI, NDH013EI, NDH016EI, NDH017EI, NDH026EI, NDH027EI, NDH028EI, NDH030EI, NDH031EI, NDH034EI, NDHR080001, NDHR080002, NDHR080003, NDHR080004, NDHR080007, NDHR080012, NDHR080021, NDHR090003, NDHR090012, and NDHR090013. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The four screws that secure the patient table top to the carrier pulled out and caused the table top to detach. This issue may affect NeuViz Dual and NeuViz 16 scanners.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Philips And Neusoft Medical Systems Co., Ltd.
- Distribuidor
- Nationwide Distribution including North Carolina, Ohio, Texas, and Puerto Rico.
- Lote
- NeuViz 16 system Serial numbers: N16E090002, N16E090003, N16E090004, N16E090007, N16E090008, N16E090010, N16E100015, N16E100017, N16E100023, N16E110016, N16E110023, N16E110026, N16E110027, N16E110040, N16E110053, N16E120016, N16E120022, N16E120020, N16E120023, N16E120024. Neuviz Dual Serial numbers: 400562, 400563, 400626, NDH009EI, NDH011EI, NDH012EI, NDH013EI, NDH016EI, NDH017EI, NDH026EI, NDH027EI, NDH028EI, NDH030EI, NDH031EI, NDH034EI, NDHR080001, NDHR080002, NDHR080003, NDHR080004, NDHR080007, NDHR080012, NDHR080021, NDHR090003, NDHR090012, and NDHR090013.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- China
Retiros relacionados
- FDANeuViz 16 Multi-Slice CT Scanner System, Part number (PN): 989605858501, Model: NeuViz 16. This is a computed1 de mayo de 2013
- FDANeuViz Dual series CT Scanner System, Part number (PN): 989605651321. This is a computed X-ray, Tomography1 de mayo de 2013
- FDASiemens syngo.plaza Radiological Image Processing System Usage: Radiological Image Processing System12 de diciembre de 2012
- FDABiOM Rechargeable Li Ion 22.2 V Battery and BiOM Lithium Battery Charger sold as an accessory to the BiOM12 de diciembre de 2012
- FDACustom Procedural Trays: Merit's Custom Total Hip Pack, K12T-05537, Merit's Total Knee Pack, K12T05538 A12 de diciembre de 2012
- FDANeuViz 16 Multi-Slice CT Scanner System is a computed Tomography X-ray system, and has a Part number1 de agosto de 2012
Preguntas sobre este retiro
¿NeuViz 16 Multi-Sliced CT Scanner System, part number 989605858501. NeuViz Dual series CT Scanner System, part number 989605651321 The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0466-2013 el 12 de diciembre de 2012. Peligro reportado: The four screws that secure the patient table top to the carrier pulled out and caused the table top to detach. This issue may affect NeuViz Dual and NeuViz 16 scanners.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter