Brand Name: HeartMate Product Name: HeartMate 3 Controller (Standalone) Model/Catalog Number: 106531US, HeartMate 3 Controller, US; 106531INT, HeartMate 3 Controller, OUS; 106531LF2, HeartMate 3 Low Flow 2.0 Controller Software Version: N/A Product Description: The HeartMate 3" System Controller acts as the central power and communication hub for the HeartMate 3 LVAS. It passes power from the Power Module, the Mobile Power Unit, Lithium-Ion Batteries, or its own integrated emergency backup supply, down to the LVAD via the Driveline. The HeartMate 3 System Controller constantly monitors system performance through communication with the implanted LVAD and Controller internal measurements and alerts the user to any alarm conditions by activating membrane panel LEDs and integrated audio annunciators. Further information on alarm conditions as well as system status can be attained by the user from the front panel LCD on the System Controller. When connected to a HeartMate Touch" Communication System, the System Controller sends information regarding the System Controller and Pump Status once per second to provide additional information to the user. This link also allows the clinician to set new patient operating parameters (e.g. pump speed) and provides a link for downloading trend and/or event recorder data. Component: Yes, HeartMate 3 LVAS
Publicado el 26 de noviembre de 2025 · Inicio del retiro: 9 de octubre de 2025
Resumen
FDA registra el retiro Z-0506-2026 de dispositivos médicos de Clase II, publicado el 26 de noviembre de 2025 e iniciado el 9 de octubre de 2025. Empresa responsable: Thoratec LLC; Estado: En curso; Cantidad reportada: 35376; Lote: Lot Code: All units of the following models are impacted: Model 106531US UDI: 05415067039446 Model 106531INT UDI: 05415067038982 Model 106531LF2 UDI: 05415067039422 Note: Product bracketing is determined based on possible use of the controller. Since the backup battery has a 3 year life cycle, it can be replaced (per the IFU) and the controller will continue to be used. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Thoratec LLC
- Distribuidor
- Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Algeria, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Lithuania, Macedonia, Malaysia, Martinique, Montenegro, Morocco, Netherlands, New Zealand, Norway, Oman, Palestine, Peru, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, UK, United Arab Emirates, Uzbekistan.
- Lote
- Lot Code: All units of the following models are impacted: Model 106531US UDI: 05415067039446 Model 106531INT UDI: 05415067038982 Model 106531LF2 UDI: 05415067039422 Note: Product bracketing is determined based on possible use of the controller. Since the backup battery has a 3 year life cycle, it can be replaced (per the IFU) and the controller will continue to be used.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDABrand Name: HeartMate Product Name: HM 11V Li-ion Backup Battery, ABT Model/Catalog Number: 106128 Software24 de junio de 2026
- FDABrand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number:24 de junio de 2026
- FDABrand Name: HeartMate Product Name: HeartMate 3" System Controller Model/Catalog Number: 106531US Software24 de junio de 2026
- FDABD Alaris infusion Pump Module 8100, REF: 8100; KIT, BEZEL ASSY 8100BD, REF: 49000437; KIT ASSY ONE PIECE26 de noviembre de 2025
- FDAMEDLINE convenience kits labeled as: 1) CENTRAL LINE PACK, REF DYNJ00281P; 2) FREE FLAP PACK, REF DYNJ68213C.26 de noviembre de 2025
- FDAHeartMate 3 System Controllers provided within the following HeartMate 3 Left Ventricular Assist System (LVAS)21 de agosto de 2024
Preguntas sobre este retiro
¿Brand Name: HeartMate Product Name: HeartMate 3 Controller (Standalone) Model/Catalog Number: 106531US, HeartMate 3 Controller, US; 106531INT, HeartMate 3 Controller, OUS; 106531LF2, HeartMate 3 Low Flow 2.0 Controller Software Version: N/A Product Description: The HeartMate 3" System Controller acts as the central power and communication hub for the HeartMate 3 LVAS. It passes power from the Power Module, the Mobile Power Unit, Lithium-Ion Batteries, or its own integrated emergency backup supply, down to the LVAD via the Driveline. The HeartMate 3 System Controller constantly monitors system performance through communication with the implanted LVAD and Controller internal measurements and alerts the user to any alarm conditions by activating membrane panel LEDs and integrated audio annunciators. Further information on alarm conditions as well as system status can be attained by the user from the front panel LCD on the System Controller. When connected to a HeartMate Touch" Communication System, the System Controller sends information regarding the System Controller and Pump Status once per second to provide additional information to the user. This link also allows the clinician to set new patient operating parameters (e.g. pump speed) and provides a link for downloading trend and/or event recorder data. Component: Yes, HeartMate 3 LVAS fue retirado del mercado?
Sí. FDA publicó el retiro Z-0506-2026 el 26 de noviembre de 2025. Peligro reportado: Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter