Tracheostomy care kits: (1) Medline Tracheostomy Clean and Care Tray with 14 Fr Suction Catheter, REF DYND40580, Sterile; (2) Medline Tracheostomy Clean and Care Tray, REF DYND4061030, Sterile; (3) Medline Basic Tracheostomy Care Kit, REF CC3T4691A, Sterile; (4) Medline Tracheostomy Care Kit with Glove, REF CC4681A, Sterile; (5) Medline Mini Tracheostomy Care Kit, REF CC3T3030A, Sterile; and (6) Medline Mini Tracheostomy Care Kit, REF DYND40622, Non-sterile.
Publicado el 3 de enero de 2024 · Inicio del retiro: 16 de octubre de 2023
Resumen
FDA registra el retiro Z-0561-2024 de dispositivos médicos de Clase I, publicado el 3 de enero de 2024 e iniciado el 16 de octubre de 2023. Empresa responsable: MEDLINE INDUSTRIES, LP - Northfield; Estado: En curso; Cantidad reportada: 620,735 kits; Lote: (1) DYND40580 - UDI 10080196032761, lot numbers: 6052203006, 6052203022, 6052204005, 6052204010, 6052205024, 6052206003, 6052206010, 6052206011, 6052206020, 6052206021, 6052207009, 6052207013, 6052207014, 6052207015,6052208021, 6052208022, 6052209009, 6052209025, 6052209026, 6052212023, 6052301031, 6052303011, 6052303013, 6052303014, 6052304020, 6052305007, 6052305019, 6052306004, 6052306014, 6052306015, 6052305026, 6052306026, 6052306032, 6052307006, 6052307011, and 6052307012. (2) DYND4061030 - UDI 10080196032877, lot numbers: 6052209025, 6052209026, 6052210016, 6052210031, 6052210032, 6052211034, 6052212009, 6052212010, 6052212021, 6052212022, 6052212015, 6052301017, 6052306014, 6052306015, 6052307011, 6052307012, 6052307022, 6052307023, 6052307026, 6052308001, 6052308003, 6052309001, 6052309003, and 6052309002. (3) CC3T4691A - UDI 10080196503391, lot numbers: 6052302012, 6052302013, 6052306004, 6052306014, 6052306015, 6052306017, 6052307012, 6052307022, 6052307023, 6052308002, and 6052308003. (4) CC4681A - UDI 10080196503407, lot numbers: 6052301017, 6052302012, 6052302013, 6052302028, 6052302029, 6052304024, 6052305020, 6052306004, 6052306014, 6052306015, 6052307011, and 6052307012. (5) CC3T3030A - UDI 10080196503384, lot numbers: 6052204010, 6052204011, 6052205021, 6052205022, 6052205023, 6052205024, 6052205028, 6052206003, 6052206010, 6052206011, 6052210015, 6052210016, 6052210031, 6052210032, 6052212010, 6052212023, 6052301015, 6052301016, 6052301017, 6052301029, 6052302012, 6052302013, 6052303011, 6052303012, 6052303013, 6052303014, 6052304019, 6052304020, 6052305007, 6052305020, 6052306004, 6052306014, 6052306015, 6052307011, 6052307012, 6052307022, 6052307023, 6052308001, 6052308002, and 6052308003. (6) DYND40622 - UDI 10080196032914, lot numbers: 6052203006, 6052204005, 6052204010, 6052205023, 6052205028, 6052206003, 6052206010, 6052207009, 6052207030, 6052208009, 6052208021, 6052208040, 6052209027, 6052209025, 6052210008, 6052210031, 6052210032, 6052212009, 6052212021, 6052212023, 6052301016, 6052302013, 6052302028, 6052303012, 6052303014, 6052305018, 6052305019, 6052305020, 6052306004, 6052306014, 6052306015, 6052306017, and 6052307022. La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The component tracheostomy brush has a sharp edge at the tip that can potentially puncture tubing during use and/or cause user injury. Additionally, there have been incidents of the tracheostomy brush bristles detaching prior to use in the packaging and/or during use.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- MEDLINE INDUSTRIES, LP - Northfield
- Distribuidor
- US Nationwide distribution including Puerto Rico. OUS (Foreign) distribution was made to Canada, Panama, and United Arab Emirates.
- Lote
- (1) DYND40580 - UDI 10080196032761, lot numbers: 6052203006, 6052203022, 6052204005, 6052204010, 6052205024, 6052206003, 6052206010, 6052206011, 6052206020, 6052206021, 6052207009, 6052207013, 6052207014, 6052207015,6052208021, 6052208022, 6052209009, 6052209025, 6052209026, 6052212023, 6052301031, 6052303011, 6052303013, 6052303014, 6052304020, 6052305007, 6052305019, 6052306004, 6052306014, 6052306015, 6052305026, 6052306026, 6052306032, 6052307006, 6052307011, and 6052307012. (2) DYND4061030 - UDI 10080196032877, lot numbers: 6052209025, 6052209026, 6052210016, 6052210031, 6052210032, 6052211034, 6052212009, 6052212010, 6052212021, 6052212022, 6052212015, 6052301017, 6052306014, 6052306015, 6052307011, 6052307012, 6052307022, 6052307023, 6052307026, 6052308001, 6052308003, 6052309001, 6052309003, and 6052309002. (3) CC3T4691A - UDI 10080196503391, lot numbers: 6052302012, 6052302013, 6052306004, 6052306014, 6052306015, 6052306017, 6052307012, 6052307022, 6052307023, 6052308002, and 6052308003. (4) CC4681A - UDI 10080196503407, lot numbers: 6052301017, 6052302012, 6052302013, 6052302028, 6052302029, 6052304024, 6052305020, 6052306004, 6052306014, 6052306015, 6052307011, and 6052307012. (5) CC3T3030A - UDI 10080196503384, lot numbers: 6052204010, 6052204011, 6052205021, 6052205022, 6052205023, 6052205024, 6052205028, 6052206003, 6052206010, 6052206011, 6052210015, 6052210016, 6052210031, 6052210032, 6052212010, 6052212023, 6052301015, 6052301016, 6052301017, 6052301029, 6052302012, 6052302013, 6052303011, 6052303012, 6052303013, 6052303014, 6052304019, 6052304020, 6052305007, 6052305020, 6052306004, 6052306014, 6052306015, 6052307011, 6052307012, 6052307022, 6052307023, 6052308001, 6052308002, and 6052308003. (6) DYND40622 - UDI 10080196032914, lot numbers: 6052203006, 6052204005, 6052204010, 6052205023, 6052205028, 6052206003, 6052206010, 6052207009, 6052207030, 6052208009, 6052208021, 6052208040, 6052209027, 6052209025, 6052210008, 6052210031, 6052210032, 6052212009, 6052212021, 6052212023, 6052301016, 6052302013, 6052302028, 6052303012, 6052303014, 6052305018, 6052305019, 6052305020, 6052306004, 6052306014, 6052306015, 6052306017, and 6052307022.
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDACURAD Alcohol Prep Pads, Sterile, Medium, 2-Ply, Contents: 5 boxes per Carton, 30 Boxes per Case, Single Use3 de septiembre de 2025
- FDAReliOn, Sterile Alcohol Swabs, Skin Cleanser, 200 Swabs, Distributed by, Walmart, Inc., Bentonville, AR 727163 de septiembre de 2025
- FDAGood Neighbor Pharmacy, Alcohol Prep Pads, Distributed By AmeriSource Bergen, 1 West First Avenue,3 de septiembre de 2025
- FDAMedline Kits containing BD SureStep Foley Tray Systems: SKU/Description: CDS984076J GYN ROBOTIC DYNJG001004A3 de septiembre de 2025
- FDABowl & Basket Zero Calorie with Sucralose Sweetener 100 count, UPC 0-41190-06855-5, 12 retail cartons per3 de enero de 2024
- FDA(1) Nursing Skills kit, SKU #EDUC05038B, Case UPC 40195327019878, Non-sterile, For Training Purposes Only, Not3 de enero de 2024
Preguntas sobre este retiro
¿Tracheostomy care kits: (1) Medline Tracheostomy Clean and Care Tray with 14 Fr Suction Catheter, REF DYND40580, Sterile; (2) Medline Tracheostomy Clean and Care Tray, REF DYND4061030, Sterile; (3) Medline Basic Tracheostomy Care Kit, REF CC3T4691A, Sterile; (4) Medline Tracheostomy Care Kit with Glove, REF CC4681A, Sterile; (5) Medline Mini Tracheostomy Care Kit, REF CC3T3030A, Sterile; and (6) Medline Mini Tracheostomy Care Kit, REF DYND40622, Non-sterile. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0561-2024 el 3 de enero de 2024. Peligro reportado: The component tracheostomy brush has a sharp edge at the tip that can potentially puncture tubing during use and/or cause user injury. Additionally, there have been incidents of the tracheostomy brush bristles detaching prior to use in the packaging and/or during use.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter