Invacare Perfecto 2 V Oxygen Concentrator The finished device is packaged as a single unit. The service part contains a single pair of sieve beds; one pair of sieve beds is used in a single finished device. Intended to provide supplemental oxygen to patients with respiratory disorders, by separating nitrogen from room air, by way of a molecular sieve.
Publicado el 17 de diciembre de 2014 · Inicio del retiro: 23 de septiembre de 2014
Resumen
FDA registra el retiro Z-0577-2015 de dispositivos médicos de Clase II, publicado el 17 de diciembre de 2014 e iniciado el 23 de septiembre de 2014. Empresa responsable: Invacare Corporation; Estado: Finalizado; Cantidad reportada: 5,963; Lote: The defective sieve bed subassembly manufactured from 7/31/2014 to 8/19/2014 impacts the finished device- IRC5PO2V and Service Part Number-1165099. The serial number range is as follows for the finished device: 14GF053772-14HF032124 (Sequential serial numbers). La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
It was determined that a defective sieve bed subassembly enabled zeolite sieve dust from the sieve beds to migrate into the valves of the manifold assembly and muffler causing a reduction in oxygen concentration.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Invacare Corporation
- Distribuidor
- Nationwide Distribution -- FL, NJ, GA, RI, PA, MI, CA, MA, IN, WA, CT, VA, NC, MO, AZ, CO, MS, MI, WI, TN, NH, IL, LA, OK, HI, KY, NE, SC, TX, IA, IN, and MS.
- Lote
- The defective sieve bed subassembly manufactured from 7/31/2014 to 8/19/2014 impacts the finished device- IRC5PO2V and Service Part Number-1165099. The serial number range is as follows for the finished device: 14GF053772-14HF032124 (Sequential serial numbers)
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAInvacare Homecare Series Bed and Invacare G-Series Beds Containing Component Model Numbers: (1) G50 (2) G5313 de marzo de 2024
- FDAAssist Bed Rail - 66327 de junio de 2017
- FDAJunction Box - 1104343 and 119356015 de febrero de 2017
- FDABariatric Bed - BARPKG-IVC 163315 de febrero de 2017
- FDASyngo.plaza, Syngo.plaza VB10A, and syngo Imaging XS : Product Usage: Syngo.plaza is a Picture Archiving and17 de diciembre de 2014
- FDAOxoid Legionella BCYE Growth Supplement packaged in packs of 10-100ml vials. A growth supplement for the17 de diciembre de 2014
Preguntas sobre este retiro
¿Invacare Perfecto 2 V Oxygen Concentrator The finished device is packaged as a single unit. The service part contains a single pair of sieve beds; one pair of sieve beds is used in a single finished device. Intended to provide supplemental oxygen to patients with respiratory disorders, by separating nitrogen from room air, by way of a molecular sieve. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0577-2015 el 17 de diciembre de 2014. Peligro reportado: It was determined that a defective sieve bed subassembly enabled zeolite sieve dust from the sieve beds to migrate into the valves of the manifold assembly and muffler causing a reduction in oxygen concentration.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter