PKG, BIPOLAR FORCEPS, FENESTRATED, 33CM, P/N 0250080123 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
Publicado el 24 de diciembre de 2014 · Inicio del retiro: 17 de noviembre de 2014
Resumen
FDA registra el retiro Z-0596-2015 de dispositivos médicos de Clase II, publicado el 24 de diciembre de 2014 e iniciado el 17 de noviembre de 2014. Empresa responsable: Stryker Endoscopy; Estado: Finalizado; Cantidad reportada: 180,573 units total; Lote: 06CON06113 0816243/0817410 0913221/0817442 0913222/0817442 1015164/1014813 1015164/1015778 1016405/1016061 1016405/1016400 1016405/1016401 1113335/1113339 1213126/1114543 1213127/1114543 1213127/1213123 1214001/1313047 1313947/1313946 0817389/0817444 0913565/0913326 0915791/0915061 0915791/1013608 0915894/0915061 0915895/0915061 1014635/1015778 1015165/1013609 1015165/1015778 1015773/1015778 1015774/1015778 1015862/1016061 1016124/1016061 1016125/1016062 1016128/1016061 1016406/1016400 1016408/1016400 1016408/1016826 1016562/1016400 1113333/1016063 1113333/1113018 1113333/1113136 1113333/1113339 1115969/1115651 1116112/1115651 1116113/1115651 1116114/1115651 1116115/1115651 1116256/1115651 1213128/1213731 1213973/1213978 1213974/1213731 1213974/1213978. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Stryker Endoscopy
- Distribuidor
- Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
- Lote
- 06CON06113 0816243/0817410 0913221/0817442 0913222/0817442 1015164/1014813 1015164/1015778 1016405/1016061 1016405/1016400 1016405/1016401 1113335/1113339 1213126/1114543 1213127/1114543 1213127/1213123 1214001/1313047 1313947/1313946 0817389/0817444 0913565/0913326 0915791/0915061 0915791/1013608 0915894/0915061 0915895/0915061 1014635/1015778 1015165/1013609 1015165/1015778 1015773/1015778 1015774/1015778 1015862/1016061 1016124/1016061 1016125/1016062 1016128/1016061 1016406/1016400 1016408/1016400 1016408/1016826 1016562/1016400 1113333/1016063 1113333/1113018 1113333/1113136 1113333/1113339 1115969/1115651 1116112/1115651 1116113/1115651 1116114/1115651 1116115/1115651 1116256/1115651 1213128/1213731 1213973/1213978 1213974/1213731 1213974/1213978
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAPKG, ENDO METZENBAUM SCISSORS, CURVED, P/N 0250080267. Laparoscopic Manual Instruments are intended for24 de diciembre de 2014
- FDAPKG, ALLIGATOR FORCEPS, P/N 0250080316. Laparoscopic Manual Instruments are intended for cutting, holding,24 de diciembre de 2014
- FDAPKG, 5MM X 33CM, INSERT, FENESTRATED GRASPER, P/N 0250080755 Laparoscopic Manual Instruments are intended for24 de diciembre de 2014
- FDAPKG, BIPOLAR FORCEPS, MICRO TIP, 33CM, P/N 0250080122 Laparoscopic Manual Instruments are intended for24 de diciembre de 2014
- FDAPKG, OP-BIOPSY FORCEPS, SPOON JAW, P/N 0250080007 Laparoscopic Manual Instruments are intended for cutting,24 de diciembre de 2014
- FDAPKG, BIPOLAR FORCEPS, PADDLE, 33CM, P/N 0250080121 Laparoscopic Manual Instruments are intended for cutting,24 de diciembre de 2014
Preguntas sobre este retiro
¿PKG, BIPOLAR FORCEPS, FENESTRATED, 33CM, P/N 0250080123 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0596-2015 el 24 de diciembre de 2014. Peligro reportado: The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter