Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0691-2025

Brand Name: K-Systems Product Name: R65 Trolley Model/Catalog Number: K24012 Software Version: N/A Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes and embryos), at closely regulated and monitored temperatures, during assisted reproduction examination and processing procedures. It may support a more comprehensive growth environment [e.g., temperature, carbon dioxide (CO2) and oxygen (O2) concentration]; it typically has adjustment and monitoring controls. Component: N/A

Publicado el 25 de diciembre de 2024 · Inicio del retiro: 15 de noviembre de 2024

Resumen

FDA registra el retiro Z-0691-2025 de dispositivos médicos de Clase II, publicado el 25 de diciembre de 2024 e iniciado el 15 de noviembre de 2024. Empresa responsable: CooperSurgical, Inc.; Estado: En curso; Cantidad reportada: 43 units; Lote: Version or Model: K24012; Primary DI Number: 00888937025026; Serial Numbers: 2234KH01, 2234KH02, 2234KH03, 2234KH04, 2234KH05, 2234KH06, 2234KH07, 2234KH08, 2234KH09, 2234KH10, 2235KH41, 2235KH42, 2235KH43, 2235KH44, 2235KH45, 2235KH46, 2235KH47, 2235KH48, 2235KH49, 2235KH50, 2235KH51, 2235KH52, 2235KH53, 2235KH54, 2235KH55, 2235KH56, 2235KH57, 2235KH58, 2235KH59, 2235KH60, 2243KH01, 2243KH02, 2243KH03, 2243KH04, 2243KH05, 2243KH06, 2243KH07, 2243KH08, 2243KH09, 2243KH10, 2243KH11, 2243KH12, 2243KH13, 2243KH14, 2243KH15, 2243KH16, 2243KH17, 2243KH18, 2243KH19, 2243KH20, 2307KH01, 2307KH02, 2307KH03, 2307KH04, 2307KH05, 2307KH06, 2307KH07, 2307KH08, 2307KH09, 2307KH10, 2320KH11, 2320KH12, 2320KH13, 2320KH14, 2320KH15, 2320KH16, 2320KH17, 2320KH18, 2320KH19, 2320KH20, 2337KH11, 2337KH12, 2337KH13, 2337KH14, 2337KH15, 2337KH16, 2337KH17, 2337KH18, 2337KH19, 2337KH20, 2340KH01, 2340KH02, 2340KH03, 2340KH04, 2340KH05,. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately.

Qué hacer

Voluntary: Firm initiated · Telephone

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
CooperSurgical, Inc.
Distribuidor
States: NC, NJ, TX, VA, IL, NY, CA, PA Please note: The countries below, excluding Iraq, all consider these devices to be "lab equipment", not medical devices and are not subject to classification or recall guidelines. Countries: AT, AU, BE, BH, BR, CA, CH, CN, CZ, DE, EG, ES, ET, FR, GB, GH, HU, ID, IE, IL, IN, IQ, IT, JO, JP, KR, KW, LB, MA, MK, MT, NL, NO, PE, PH, PK, PT, QA, RO, RS, SA, SD, SE, SO,TR, TW, TZ, US, ZA
Lote
Version or Model: K24012; Primary DI Number: 00888937025026; Serial Numbers: 2234KH01, 2234KH02, 2234KH03, 2234KH04, 2234KH05, 2234KH06, 2234KH07, 2234KH08, 2234KH09, 2234KH10, 2235KH41, 2235KH42, 2235KH43, 2235KH44, 2235KH45, 2235KH46, 2235KH47, 2235KH48, 2235KH49, 2235KH50, 2235KH51, 2235KH52, 2235KH53, 2235KH54, 2235KH55, 2235KH56, 2235KH57, 2235KH58, 2235KH59, 2235KH60, 2243KH01, 2243KH02, 2243KH03, 2243KH04, 2243KH05, 2243KH06, 2243KH07, 2243KH08, 2243KH09, 2243KH10, 2243KH11, 2243KH12, 2243KH13, 2243KH14, 2243KH15, 2243KH16, 2243KH17, 2243KH18, 2243KH19, 2243KH20, 2307KH01, 2307KH02, 2307KH03, 2307KH04, 2307KH05, 2307KH06, 2307KH07, 2307KH08, 2307KH09, 2307KH10, 2320KH11, 2320KH12, 2320KH13, 2320KH14, 2320KH15, 2320KH16, 2320KH17, 2320KH18, 2320KH19, 2320KH20, 2337KH11, 2337KH12, 2337KH13, 2337KH14, 2337KH15, 2337KH16, 2337KH17, 2337KH18, 2337KH19, 2337KH20, 2340KH01, 2340KH02, 2340KH03, 2340KH04, 2340KH05,
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Brand Name: K-Systems Product Name: R65 Trolley Model/Catalog Number: K24012 Software Version: N/A Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes and embryos), at closely regulated and monitored temperatures, during assisted reproduction examination and processing procedures. It may support a more comprehensive growth environment [e.g., temperature, carbon dioxide (CO2) and oxygen (O2) concentration]; it typically has adjustment and monitoring controls. Component: N/A fue retirado del mercado?

Sí. FDA publicó el retiro Z-0691-2025 el 25 de diciembre de 2024. Peligro reportado: The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Telephone