Becton, Dickinson and Company, BBL Crystal Enteric/Nonfermenter ID Kit, Catalog # 24500, cartons of 20 kits, In-vitro diagnostic reagents for microbiological testing. The BBL Crystal Enteric/Nonfermentor (E/NF) identification (ID) System is for the identification of aerobic gram-negative bacteria that belong to the family Enterobacteriaceae as well as some of the more frequently isolated glucose fermenting and nonfermenting gram-negative bacilli.
Publicado el 30 de enero de 2013 · Inicio del retiro: 13 de diciembre de 2012
Resumen
FDA registra el retiro Z-0699-2013 de dispositivos médicos de Clase II, publicado el 30 de enero de 2013 e iniciado el 13 de diciembre de 2012. Empresa responsable: Becton Dickinson & Co.; Estado: Finalizado; Cantidad reportada: 67,820 units; Lote: Lot Number/ Exp Date 2086315 February 20, 2013 2101435 February 20, 2013 2104316 February 27, 2013 2129460 February 27, 2013 2129487 March 13, 2013 2163211 February 27, 2013 2163213 March 20, 2013 2163215 March 27, 2013. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
In-vitro diagnostic reagents for microbiological testing may be reversed in well positions in the test kit and when utilized may lead to bacteria misidentification.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Becton Dickinson & Co.
- Distribuidor
- Worldwide Distribution-USA including the states of MA, TN, KY, IN, GA, NC, TX, NY, FL, AZ, WA, NJ, MI, and CA, and the countries of Uruguay, Taiwan, Singapore, Australia, South Korea, Mexico, Hong Kong, Guatemala, Colombia, Chile, Canada, India, Japan, Brazil, China and Belgium.
- Lote
- Lot Number/ Exp Date 2086315 February 20, 2013 2101435 February 20, 2013 2104316 February 27, 2013 2129460 February 27, 2013 2129487 March 13, 2013 2163211 February 27, 2013 2163213 March 20, 2013 2163215 March 27, 2013
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDABD Phoenix Panel NMIC-312. Catalog Number: 449082.21 de octubre de 2026
- FDABD Phoenix Panel NMIC-306. Catalog Number: 449292.21 de octubre de 2026
- FDABD GasPak EZ CO2 Pouch System. Catalog Number: 260684. for microbiologic use6 de mayo de 2026
- FDABD GasPak EZ Campy Pouch System. Catalog Number: 260685. for microbiologic use6 de mayo de 2026
- FDAAbbott Laboratories, ARCHITECT 25-OH Vitamin D Reagent Kit, 03L52-25 (ROW) and 03L52-27 (U.S.) (1 x 10030 de enero de 2013
- FDAGreenLight HPS (High Performance System) GreenLight XPS (Xcelerated Performance System) Manufactured by AMS,30 de enero de 2013
Preguntas sobre este retiro
¿Becton, Dickinson and Company, BBL Crystal Enteric/Nonfermenter ID Kit, Catalog # 24500, cartons of 20 kits, In-vitro diagnostic reagents for microbiological testing. The BBL Crystal Enteric/Nonfermentor (E/NF) identification (ID) System is for the identification of aerobic gram-negative bacteria that belong to the family Enterobacteriaceae as well as some of the more frequently isolated glucose fermenting and nonfermenting gram-negative bacilli. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0699-2013 el 30 de enero de 2013. Peligro reportado: In-vitro diagnostic reagents for microbiological testing may be reversed in well positions in the test kit and when utilized may lead to bacteria misidentification.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter