Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0809-2025

Xhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system with patient data.

Publicado el 15 de enero de 2025 · Inicio del retiro: 25 de noviembre de 2024

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Telemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts; necessary clinical intervention could be delayed.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Spacelabs Healthcare, Inc.
Distribuidor
US: OK, MN, IA, SC, NE, MI, NY, KY, GA, MO, MT, LA, WA, SD, CO, PA, WY, TX, ID, NJ, FL, OH, MS, OR, AK, IL, AR, KS, AL, TN, WI, VA, NM, CA, IN, RI, NV, MA, NC, ME, CT, AZ, MD, PR. OUS: Kuwait, Australia, Canada, Roumanie, Greece, Philippines, Spain, Italy, Hong Kong, Switzerland, Panama, Taiwan, Netherlands, Malaysia, Poland, Mexico, United Kingdom, Saudi Arabia, France, Czech Republic
Lote
Service Manuals P/N 070-2409-07 and Prior. UDI-DI: 10841522107177, 10841522100246
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta