Puritan Bennett 840 Series Ventilator PB840 Operators Manual Addendum PN 066009A 09/02 The intended use of the oxygen sensor is to provide the oxygen concentration of the delivered gas at the patient wye (inspiratory-expiratory interface). The oxygen sensor does not control the oxygen level or the flow of gases. The intended use of the devices is as critical care ventilators that provide continuous mechanical ventilation for patients.
Publicado el 29 de enero de 2014 · Inicio del retiro: 25 de enero de 2013
Resumen
FDA registra el retiro Z-0821-2014 de dispositivos médicos de Clase II, publicado el 29 de enero de 2014 e iniciado el 25 de enero de 2013. Empresa responsable: Nellcor Puritan Bennett Inc. (dba Covidien LP); Estado: Finalizado; Cantidad reportada: 70,160; Lote: PB840 Operators Manual Addendum: PN 066009A 09/02. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Covidien is correcting labeling to clarify the operational life of the oxygen sensor (O2 sensor), PN 4-072214-00 used in the Puritan Bennett" 840, and PN, G-062010-00, used in Puritan Bennett" 740, and760 ventilator systems. The Operators Manual for PB840 states a nominal life of one year and the Addendum incorrectly states 2 years.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Nellcor Puritan Bennett Inc. (dba Covidien LP)
- Distribuidor
- Worldwide Distribution: US (nationwide) and Internationally to: Albania, Argentina, Australia, Bahrain, Belgium, Belize, Bermuda Bolivia, Bosnia and Herzegovina, Brazil, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Croatia, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Greece, Guam, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Macedonia, The former Yugoslav, Malaysia, Mexico, Moldova, Republic of, Montenegro, Morocco, Netherlands, Nicaragua, Northern Mariana Islands, Oman, Panama, Paraguay, Peru, Philippines, Poland, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Switzerland, Syria, Taiwan, Thailand, Tunisia, Turkey, Uganda, Ukrainian SSR, United Arab Emirates (UAE) United Kingdom, Uruguay, Venezuela and Yugoslavia.
- Lote
- PB840 Operators Manual Addendum: PN 066009A 09/02
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDACovidien Puritan Bennett 980 Ventilator, Rx ONLY, Made in Ireland., Covidien LLC, USA. REF 980X1ENDIUU.17 de diciembre de 2014
- FDACovidien Puritan Bennett 980 Ventilator, Rx ONLY. Suitable for service in a hospital (institutions) and5 de noviembre de 2014
- FDAVital Sync" VPMP & IM, software, V2.4. (DVD), packaged on a CD or other electronically controlled device. CDs20 de agosto de 2014
- FDAVital Sync RMS 1.0 when loaded with v2.4 software, packaged on a CD or other electronically controlled device.20 de agosto de 2014
- FDAComposix L/P with Echo PS 6" X 8" Product Code 0144680 The Composix L/P Mesh with Echo PS Positioning System29 de enero de 2014
- FDAPuritan Bennett 700 Series Ventilators (PB740 and PB760) PB700 Series Operators Manual: PN 10066984 & PN29 de enero de 2014
Preguntas sobre este retiro
¿Puritan Bennett 840 Series Ventilator PB840 Operators Manual Addendum PN 066009A 09/02 The intended use of the oxygen sensor is to provide the oxygen concentration of the delivered gas at the patient wye (inspiratory-expiratory interface). The oxygen sensor does not control the oxygen level or the flow of gases. The intended use of the devices is as critical care ventilators that provide continuous mechanical ventilation for patients. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0821-2014 el 29 de enero de 2014. Peligro reportado: Covidien is correcting labeling to clarify the operational life of the oxygen sensor (O2 sensor), PN 4-072214-00 used in the Puritan Bennett" 840, and PN, G-062010-00, used in Puritan Bennett" 740, and760 ventilator systems. The Operators Manual for PB840 states a nominal life of one year and the Addendum incorrectly states 2 years.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter