Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation.

Publicado el 2 de marzo de 2016 · Inicio del retiro: 4 de febrero de 2016

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware version 1.50 released in November 2015 (recall Z-0436-2016). The batteries were prematurely discharged, even though an adequate battery charge status was displayed on the device.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Draeger Medical, Inc.
Distribuidor
US Nationwide Distribution including Puerto Rico.
Lote
Catalog #s 8416400 and 8417400
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States