Custom procedural convenience kits and trays, Ophthalmic use, labeled as: a) EYE PACK, kit number LVEY27; b) Laminectomy, kit number UIOL34AG

Publicado el 7 de febrero de 2024 · Inicio del retiro: 29 de noviembre de 2023

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

Qué hacer

Voluntary: Firm initiated · E-Mail

Detalles del retiro

Fabricante
American Contract Systems, Inc.
Distribuidor
US Nationwide distribution.
Lote
a) LVEY27, UDI/DI 00191072135728, Lot Numbers: 840221, exp. 6/29/2023; 771221, exp. 8/17/2023; 715221, exp. 10/12/2023; 982231, exp. 1/18/2024; 917231, exp. 3/24/2023; b) UIOL34AG, UDI/DI 00191072191168, Lot Numbers: 912231, exp. 3/29/2024; 898231, exp. 4/12/2024
Clasificación de riesgo
Class II