Fluorescence Imaging Procedure Kit, designed for use with Firefly Fluorescence Imaging for the da Vinci Si 3000 Surgical Systems. The Intuitive Surgical Endoscopic Instrument Control System (Intuitive Surgical da Vinci Surgical System Model IS3000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.

Publicado el 31 de diciembre de 2014 · Inicio del retiro: 21 de noviembre de 2014

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The storage conditions for indocyanine green (ICG) are not on the individual Fluorescence Imaging Procedure Kit box, the six-kit box, and the shipping box. If heat were to damage the material, the result would be a loss of fluorescence.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Intuitive Surgical, Inc.
Distribuidor
Nationwide Distribution.
Lote
PN 950156-01
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States