AQURE basic sytem, FLEXLINK, REF 933-599, Version: 2.3.3 (17), CE IVD, Rx Only, UDI: (01)05700699335999 Intended use: The AQURE system is intended to let you manage analytical devices and operator profiles.

Publicado el 5 de febrero de 2020 · Inicio del retiro: 8 de agosto de 2019

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The firm has become aware that there is a potential problem relating to the blood gas and immunoassay analyzer Systems that may result in patient mix-up when connected to some third-party devices. The error may lead to serious adverse health consequences for the patient caused by patient data mix-up.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Radiometer Medical ApS
Distribuidor
Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, FL, GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI and WV; and countries of:  Australia, Austria, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Hungary, India, Italy, Japan (Tokyo-not a country but it was listed), Netherlands, Norway, Poland, Singapore, South Korea, Spain, Sweden, Switzerland, Turkey and United Kingdom.
Lote
Software version 2.3.5 and below
Clasificación de riesgo
Class II
Área (FDA)
device
País
Denmark