e.cam -emission computed tomography system Used to detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV.
Publicado el 14 de enero de 2015 · Inicio del retiro: 8 de abril de 2014
Resumen
FDA registra el retiro Z-0914-2015 de dispositivos médicos de Clase II, publicado el 14 de enero de 2015 e iniciado el 8 de abril de 2014. Empresa responsable: Siemens Medical Solutions USA, Inc.; Estado: Finalizado; Cantidad reportada: 4,462 systems; Lote: 1) E.CAM DCR Duet Head, A (e.soft); Serial Numbers: 7090 , 7501 , 7502 , 8001 , 8002 , 8003 , 8004 , 8501 , 8502 , 8506 , 8507 ; 2) E.CAM DCR Gantry Dual Head, Bl Serial Numbers: 3097 , 8502 , 8507 , 8508 , 8509 , 8510 , 8511 , 8514 , 8516 , 8521 , 8522 , 9821 , 71101 ; 3) E.CAM DCR Gantry Single Head; Serial Numbers: 5045 , 7501 , 7502 , 7503 , 7504 , 7505 , 7506 , 7507 , 7508 , 7509 , 7510 , 7511 , 7512 , 8001 , 8002 , 8003 , 8004 , 8005 , 8006 , 8007 , 8008 , 8009 , 8010 , 8011 , 8012 , 8013 , 8014 , 8015 , 8016 , 8017 , 8018 , 8019 , 8020 , 8021 , 8022 , 8023 , 8024 , 8025 , 8026 , 8027 , 8028 , 8029 , 8030 , 8031 , 8032 , 8033 , 8501 , 8502 , 8503 , 8504 , 8505 , 8506 , 8507 , 8508 , 8509 , 8510 , 8511 , 8512 , 8513 , 8514 , 8516 , 8517 , 8518 , 8519 , 8520 , 8521 , 8522 , 8523 , 8524 , 8525 , 8526 , 8527 , 8528 , 8530 , 8532 , 8533 , 8534 , 8535 , 8536 , 8537 , 8538 , 8539 , 8540 , 8541 , 8542 , 8543 , 8544 , 8545 , 8546 , 8547 , 8548 , 8549 , 8550 , 8551 , 8552 , 8553 , 8554 , 8555 , 8556 ; 4) E.CAM DCR Gantry Single Head, B; Serial Numbers: 2011 , 5504 , 7510 , 8501 , 8502 , 8503 , 8505 , 8507 , 8508 , 8509 , 8511 , 8512 , 8513 , 8514 , 8516 , 8517 , 8518 , 8519 , 8520 , 8521 , 8501* ; 5) E.CAM Duet Head (e.soft); Serial Numbers: 2 , 3 , 6001 , 6003 , 6005 , 6006 , 6007 , 6009 , 6010 , 7001 , 7004 , 7005 , 7006 , 7502 , 7503 , 7505 , 7507 , 7508 ; 6) E.CAM extended Gantry dual head v3; Serial Numbers: 1001 , 5039 , 5161 , 5501 , 5504 , 5506 , 5507 , 5508 , 5509 , 5510 , 5511 , 5512 , 5513 , 5514 , 5515 , 5516 , 5517 , 5519 , 5520 , 5521 , 5522 , 5524 , 5525 , 5526 , 5528 , 5529 , 5530 , 5532 , 5533 , 5534 , 5535 , 5536 , 5538 , 5539 , 5540 , 5541 , 5542 , 5543 , 5544 , 5545 , 5546 , 5547 , 5548 , 5549 , 5550 , 5553 , 5554 , 5555 , 5556 , 5557 , 5558 , 5560 , 5561 , 5563 , 5564 , 5565 , 5566 , 5567 , 5568 , 5570 , 5571 , 5572 , 5576 , 5578 , 5579 , 5580 , 5582 , 5583 , 5584 , 5585 , 5586 , 5587 , 5588 , 5589 , 5590 , 5591 , 5592 , 5593 , 5594 , 5595 , 5596 , 5597 , 5598 , 5599 , 5600 , 5601 , 5602 , 5603 , 5604 , 5605 , 5606 , 5607 , 5608 , 5609 , 5610 , 5611 , 5612 , 5613 , 5614 , 5615 , 5616 , 5617 , 5618 , 5619 , 5620 , 5621 , 5622 , 5623 , 5624 , 5625 , 5626 , 5628 , 5629 , 5631 , 5632 , 5633 , 5634 , 5635 , 5636 , 5637 , 5638 , 5639 , 5640 , 5641 , 5642 , 5643 , 5644 , 5645 , 5646 , 5647 , 5649 , 5650 , 5652 , 5655 , 5658 , 7002 , 7003 , 7004 , 7005 , 7006 , 7007 , 7008 , 7009 , 7010 , 7011 , 7012 , 7013 , 7014 , 7016 , 7017 , 7018 , 7019 , 7020 , 7021 , 7022 , 7023 , 7024 , 7025 , 7026 , 7027 , 7028 , 7029 , 7030 , 7031 , 7032 , 7033 , 7034 , 7036 , 7037 , 7038 , 7039 , 7040 , 7041 , 7042 , 7043 , 7044 , 7045 , 7046 , 7047 , 7048 , 7049 , 7050 , 7051 , 7052 , 7053 , 7054 , 7055 , 7056 , 7057 , 7058 , 7059 , 7060 , 7061 , 7062 , 7063 , 7065 , 7067 , 7068 , 7069 , 7070 , 7071 , 7072 , 7073 , 7074 , 7075 , 7076 , 7077 , 7078 , 7079 , 7080 , 7081 , 7082 , 7083 , 7084 , 7085 , 7086 , 7088 , 7090 , 7091 , 7092 , 7093 , 7094 , 7095 , 7096 , 7097 , 7097 , 7098 , 7099 , 7101 , 7103 , 7104 , 7105 , 7109 , 7110 , 7111 , 7112 , 7114 , 7115 , 7116 , 7117 , 7118 , 7119 , 7120 , 7121 , 7122 , 7123 , 7124 , 7125 , 7126 , 7127 , 7129 , 7130 , 7178 , 7179 , 7501 , 7502 , 7504 , 7505 , 7506 , 7507 , 7508 , 7509 , 7510 , 7511 , 7512 , 7513 , 7514 , 7515 , 7516 , 7518 , 7521 , 7522 , 7523 , 7524 , 7526 , 7528 , 7529 , 7530 , 7531 , 7532 , 7533 , 7534 , 7535 , 7536 , 7537 , 7538 , 7539 , 7540 , 7541 , 7542 , 7543 , 7544 , 7545 , 7546 , 7549 , 7550 , 7552 , 7553 , 7554 , 7556 , . La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
There is a potential for patient finger injury on the e.cam systems.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Siemens Medical Solutions USA, Inc.
- Distribuidor
- Worldwide Distribution: US (nationwide) including states of: AK, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, MA, ME, MI, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WY and Puerto Rico; and countries of: Algeria , Angola , Argentina , Australia , Austria , Azerbaijan , Bahrain , Bangladesh , Belarus , Belgium , Bolivia , Bosnia Herzeg. , Brazil , Bulgaria , Cameroon , Canada , Chile , Colombia , Costa Rica , Croatia , Cuba , Czech Republic , Dem. Rep. Congo , Denmark , Dominican Rep. , Ecuador , Egypt , El Salvador , Finland , France , Georgia , Germany , Ghana , Greece , Guam , Honduras , Hong Kong , India , Indonesia , Iran , Iraq , Ireland , Israel , Italy , Jamaica , Japan , Jordan , Kazakhstan , Kosovo , Kuwait , Latvia , Lebanon , Luxembourg , Macedonia , Malaysia , Mauritius , Mexico , Morocco , Myanmar , Nepal , Netherlands , New Zealand , Niger , Nigeria , Norway , Oman , P.R. China , Pakistan , Palestinian Ter , Peru , Philippines , Poland , Portugal , Qatar , Rep. of Yemen , Republic Korea , Romania , Russian Fed. , Saudi Arabia , Serbia , Singapore , Slovakia , Slovenia , South Africa , Spain , Sri Lanka , Sweden , Switzerland , Syria , Taiwan , Thailand , Trinidad,Tobago , Tunisia , Turkey , Uganda , Ukraine , United Arab Emirates , United Kingdom , Uruguay , Uzbekistan , and Vietnam.
- Lote
- 1) E.CAM DCR Duet Head, A (e.soft); Serial Numbers: 7090 , 7501 , 7502 , 8001 , 8002 , 8003 , 8004 , 8501 , 8502 , 8506 , 8507 ; 2) E.CAM DCR Gantry Dual Head, Bl Serial Numbers: 3097 , 8502 , 8507 , 8508 , 8509 , 8510 , 8511 , 8514 , 8516 , 8521 , 8522 , 9821 , 71101 ; 3) E.CAM DCR Gantry Single Head; Serial Numbers: 5045 , 7501 , 7502 , 7503 , 7504 , 7505 , 7506 , 7507 , 7508 , 7509 , 7510 , 7511 , 7512 , 8001 , 8002 , 8003 , 8004 , 8005 , 8006 , 8007 , 8008 , 8009 , 8010 , 8011 , 8012 , 8013 , 8014 , 8015 , 8016 , 8017 , 8018 , 8019 , 8020 , 8021 , 8022 , 8023 , 8024 , 8025 , 8026 , 8027 , 8028 , 8029 , 8030 , 8031 , 8032 , 8033 , 8501 , 8502 , 8503 , 8504 , 8505 , 8506 , 8507 , 8508 , 8509 , 8510 , 8511 , 8512 , 8513 , 8514 , 8516 , 8517 , 8518 , 8519 , 8520 , 8521 , 8522 , 8523 , 8524 , 8525 , 8526 , 8527 , 8528 , 8530 , 8532 , 8533 , 8534 , 8535 , 8536 , 8537 , 8538 , 8539 , 8540 , 8541 , 8542 , 8543 , 8544 , 8545 , 8546 , 8547 , 8548 , 8549 , 8550 , 8551 , 8552 , 8553 , 8554 , 8555 , 8556 ; 4) E.CAM DCR Gantry Single Head, B; Serial Numbers: 2011 , 5504 , 7510 , 8501 , 8502 , 8503 , 8505 , 8507 , 8508 , 8509 , 8511 , 8512 , 8513 , 8514 , 8516 , 8517 , 8518 , 8519 , 8520 , 8521 , 8501* ; 5) E.CAM Duet Head (e.soft); Serial Numbers: 2 , 3 , 6001 , 6003 , 6005 , 6006 , 6007 , 6009 , 6010 , 7001 , 7004 , 7005 , 7006 , 7502 , 7503 , 7505 , 7507 , 7508 ; 6) E.CAM extended Gantry dual head v3; Serial Numbers: 1001 , 5039 , 5161 , 5501 , 5504 , 5506 , 5507 , 5508 , 5509 , 5510 , 5511 , 5512 , 5513 , 5514 , 5515 , 5516 , 5517 , 5519 , 5520 , 5521 , 5522 , 5524 , 5525 , 5526 , 5528 , 5529 , 5530 , 5532 , 5533 , 5534 , 5535 , 5536 , 5538 , 5539 , 5540 , 5541 , 5542 , 5543 , 5544 , 5545 , 5546 , 5547 , 5548 , 5549 , 5550 , 5553 , 5554 , 5555 , 5556 , 5557 , 5558 , 5560 , 5561 , 5563 , 5564 , 5565 , 5566 , 5567 , 5568 , 5570 , 5571 , 5572 , 5576 , 5578 , 5579 , 5580 , 5582 , 5583 , 5584 , 5585 , 5586 , 5587 , 5588 , 5589 , 5590 , 5591 , 5592 , 5593 , 5594 , 5595 , 5596 , 5597 , 5598 , 5599 , 5600 , 5601 , 5602 , 5603 , 5604 , 5605 , 5606 , 5607 , 5608 , 5609 , 5610 , 5611 , 5612 , 5613 , 5614 , 5615 , 5616 , 5617 , 5618 , 5619 , 5620 , 5621 , 5622 , 5623 , 5624 , 5625 , 5626 , 5628 , 5629 , 5631 , 5632 , 5633 , 5634 , 5635 , 5636 , 5637 , 5638 , 5639 , 5640 , 5641 , 5642 , 5643 , 5644 , 5645 , 5646 , 5647 , 5649 , 5650 , 5652 , 5655 , 5658 , 7002 , 7003 , 7004 , 7005 , 7006 , 7007 , 7008 , 7009 , 7010 , 7011 , 7012 , 7013 , 7014 , 7016 , 7017 , 7018 , 7019 , 7020 , 7021 , 7022 , 7023 , 7024 , 7025 , 7026 , 7027 , 7028 , 7029 , 7030 , 7031 , 7032 , 7033 , 7034 , 7036 , 7037 , 7038 , 7039 , 7040 , 7041 , 7042 , 7043 , 7044 , 7045 , 7046 , 7047 , 7048 , 7049 , 7050 , 7051 , 7052 , 7053 , 7054 , 7055 , 7056 , 7057 , 7058 , 7059 , 7060 , 7061 , 7062 , 7063 , 7065 , 7067 , 7068 , 7069 , 7070 , 7071 , 7072 , 7073 , 7074 , 7075 , 7076 , 7077 , 7078 , 7079 , 7080 , 7081 , 7082 , 7083 , 7084 , 7085 , 7086 , 7088 , 7090 , 7091 , 7092 , 7093 , 7094 , 7095 , 7096 , 7097 , 7097 , 7098 , 7099 , 7101 , 7103 , 7104 , 7105 , 7109 , 7110 , 7111 , 7112 , 7114 , 7115 , 7116 , 7117 , 7118 , 7119 , 7120 , 7121 , 7122 , 7123 , 7124 , 7125 , 7126 , 7127 , 7129 , 7130 , 7178 , 7179 , 7501 , 7502 , 7504 , 7505 , 7506 , 7507 , 7508 , 7509 , 7510 , 7511 , 7512 , 7513 , 7514 , 7515 , 7516 , 7518 , 7521 , 7522 , 7523 , 7524 , 7526 , 7528 , 7529 , 7530 , 7531 , 7532 , 7533 , 7534 , 7535 , 7536 , 7537 , 7538 , 7539 , 7540 , 7541 , 7542 , 7543 , 7544 , 7545 , 7546 , 7549 , 7550 , 7552 , 7553 , 7554 , 7556 ,
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAACUSON Redwood 1.0 and 2.0 Diagnostic Ultrasound Systems, REF: 11503314, with software2 de octubre de 2024
- FDAACUSON Juniper 1.0, 1.5, 2.0, 2.5 Diagnostic Ultrasound Systems, REF: 11335791, with software. ACUSON Juniper2 de octubre de 2024
- FDAACUSON Maple 1.0 Diagnostic Ultrasound System, REF: 11711750, with software. ACUSON Maple Select 1.02 de octubre de 2024
- FDAACUSON Redwood 1.0 Diagnostic Ultrasound System, REF: 11503314, running software version VA10A to VA10G1 de mayo de 2024
- FDAVintage Pure Bee Flower Honey, Net Weight 16.4 oz (465g), Packaged in a Glass Jar, Product of Turkey, Imported14 de enero de 2015
- FDACable remote control JUPITER / TS7500, Material no.1601926, Serial number: to (incl.) 101465919. Patient14 de enero de 2015
Preguntas sobre este retiro
¿e.cam -emission computed tomography system Used to detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0914-2015 el 14 de enero de 2015. Peligro reportado: There is a potential for patient finger injury on the e.cam systems.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter