SmartSite Needleless Connector, Model No. 2000E The SmartSite Needleless Connectors are sterile, individually packaged connectors. The connectors are used to administer fluid and medications. The needle free connector allows the user to add medication into the primary line without the use of a needle.
Publicado el 12 de febrero de 2014 · Inicio del retiro: 22 de enero de 2014
Resumen
FDA registra el retiro Z-0925-2014 de dispositivos médicos de Clase II, publicado el 12 de febrero de 2014 e iniciado el 22 de enero de 2014. Empresa responsable: CareFusion 303, Inc.; Estado: Finalizado; Cantidad reportada: 3,500,000 units (2,136,200 units in US; 1,404,700 units foreign); Lote: Lot Numbers: 13045933 13046117 13086415 13086446 13095237 13095302 13095324 13095579 13095674 13095675 13095777 13095778 13095808 13095944 13096011 13096012 13096013 13096158 13096723 13096770 13105584 13106017 13106018 13106096 13106260 13106355 13106451 13106497 13106498 13106753 13106754 13106847 13106848 13106849 13115209 13115210 13115317 13115733 13115734 13115871 13115872 13125361 13125362. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
CareFusion is recalling the SmartSite Needlefree Connector, model number 2000E, because of connection issues. The affected SmartSite connector lots may unintentionally disconnect from a female luer, may be difficult to disconnect from a female luer, or may fail to disconnect from a female luer once attached. Leakage may also be observed if the connector disconnects from the female luer during in
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- CareFusion 303, Inc.
- Distribuidor
- Worldwide Distribution: US (nationwide) including Puerto Rico; and internationally to: Australia, Canada and New Zealand.
- Lote
- Lot Numbers: 13045933 13046117 13086415 13086446 13095237 13095302 13095324 13095579 13095674 13095675 13095777 13095778 13095808 13095944 13096011 13096012 13096013 13096158 13096723 13096770 13105584 13106017 13106018 13106096 13106260 13106355 13106451 13106497 13106498 13106753 13106754 13106847 13106848 13106849 13115209 13115210 13115317 13115733 13115734 13115871 13115872 13125361 13125362
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDABD Pyxis Anesthesia Station 4000 REF: 338 Medication cabinet10 de junio de 2026
- FDABD Pyxis MedBank Tower REF: 1145-00 Medication cabinet10 de junio de 2026
- FDABD Pyxis Mini Main REF: 349 Medication cabinet10 de junio de 2026
- FDABD Pyxis MedStation ES REF: 323 Medication cabinet10 de junio de 2026
- FDAABX PENTRA Urine Control L/H. The ABX PENTRA Urine Control L/H is a two-level (Low and High) quality control12 de febrero de 2014
- FDAThe Rotoprone therapy System is an advanced patient care system for the treatment and prevention of pulmonary12 de febrero de 2014
Preguntas sobre este retiro
¿SmartSite Needleless Connector, Model No. 2000E The SmartSite Needleless Connectors are sterile, individually packaged connectors. The connectors are used to administer fluid and medications. The needle free connector allows the user to add medication into the primary line without the use of a needle. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0925-2014 el 12 de febrero de 2014. Peligro reportado: CareFusion is recalling the SmartSite Needlefree Connector, model number 2000E, because of connection issues. The affected SmartSite connector lots may unintentionally disconnect from a female luer, may be difficult to disconnect from a female luer, or may fail to disconnect from a female luer once attached. Leakage may also be observed if the connector disconnects from the female luer during in
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter