Medtronic O-arm Imaging System Mobile X-ray system. Model No. BI-700-00027 and BI-700-00028 The 0-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-ann Imaging System is compatible with certain Image Guided Surgery Systems.
Publicado el 12 de marzo de 2014 · Inicio del retiro: 25 de julio de 2013
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Review of on-system electronic product labeling and Instructions for comply with applicable Use in response to a customer inquiry uncovered gaps in compliance FDA standards with the requirements for information to be provided to users in 21CFR 1020.30(h)
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Medtronic Navigation, Inc.
- Distribuidor
- USA (nationwide) Distribution
- Lote
- Model No. BI-700-00027 and BI-700-00028 Serial number range: 00102 through 00753 and B13080721, B13080722, B13090729, B13120763, B13120764, B13120766, B13120768, B13120769
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States