Medtronic O-arm Imaging System Mobile X-ray system. Model No. BI-700-00027 and BI-700-00028 The 0-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-ann Imaging System is compatible with certain Image Guided Surgery Systems.
Publicado el 12 de marzo de 2014 · Inicio del retiro: 25 de julio de 2013
Resumen
FDA registra el retiro Z-0973-2014 de dispositivos médicos de Clase II, publicado el 12 de marzo de 2014 e iniciado el 25 de julio de 2013. Empresa responsable: Medtronic Navigation, Inc.; Estado: Finalizado; Cantidad reportada: 449 in US; Lote: Model No. BI-700-00027 and BI-700-00028 Serial number range: 00102 through 00753 and B13080721, B13080722, B13090729, B13120763, B13120764, B13120766, B13120768, B13120769. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Review of on-system electronic product labeling and Instructions for comply with applicable Use in response to a customer inquiry uncovered gaps in compliance FDA standards with the requirements for information to be provided to users in 21CFR 1020.30(h)
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Medtronic Navigation, Inc.
- Distribuidor
- USA (nationwide) Distribution
- Lote
- Model No. BI-700-00027 and BI-700-00028 Serial number range: 00102 through 00753 and B13080721, B13080722, B13090729, B13120763, B13120764, B13120766, B13120768, B13120769
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAMedtronic Biopsy Needle Kit, REF: 9733068, Passive, Sterile EO, Rxonly28 de octubre de 2026
- FDAMedtronic NDI Passive Spheres REF: 8801075, Sterile EO, 5/Tray 12 Pk., Rxonly28 de octubre de 2026
- FDAMedtronic NDI Passive Spheres REF: 8801071, Sterile EO, 1/Tray 12 pk., Rxonly28 de octubre de 2026
- FDAMedtronic NDI Passive Spheres REF: 8801074, Sterile EO, 4/Tray 12 Pk., Rxonly28 de octubre de 2026
- FDATSX-301C Aquilion One (Vision Edition) Computed Tomography X-Ray System.12 de marzo de 2014
- FDAMcKesson Anesthesia Care Product Usage: The McKesson Anesthesia Care system is a computer based system which12 de marzo de 2014
Preguntas sobre este retiro
¿Medtronic O-arm Imaging System Mobile X-ray system. Model No. BI-700-00027 and BI-700-00028 The 0-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-ann Imaging System is compatible with certain Image Guided Surgery Systems. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0973-2014 el 12 de marzo de 2014. Peligro reportado: Review of on-system electronic product labeling and Instructions for comply with applicable Use in response to a customer inquiry uncovered gaps in compliance FDA standards with the requirements for information to be provided to users in 21CFR 1020.30(h)
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter