CareStream Dental CS 2100 X-ray system Product Usage: The expected usage (intended use) for all of these products is for dental radiographic examination and diagnosis of the teeth, jaw and oral structures.
Publicado el 8 de mayo de 2013 · Inicio del retiro: 18 de septiembre de 2012
Resumen
FDA registra el retiro Z-0999-2013 de dispositivos médicos de Clase II, publicado el 8 de mayo de 2013 e iniciado el 18 de septiembre de 2012. Empresa responsable: Trophy Sas; Estado: Finalizado; Cantidad reportada: 167 installed in the US; Lote: ABYJ011 ABYJ012 ABYJ020 through ABYJ026 ACJY522 ACYJ004 through ACYJ007 ACYJ009 through ACYJ012 ACYJ014 ACYJ015 ACYJ017 through ACYJ023 ACYJ500 through ACYJ509 ACYJ517 through ACYJ521 ACYJ523 through ACYJ531 ACYJ533 through ACYJ536 ADYJ001 through ADYJ005 ADYJ007 ADYJ008 ADYJ011 through ADYJ013 ADYJ015 through ADYJ018 ADYJ020 through ADYJ032 ADYJ034 through ADYJ037 ADYJ039 through ADYJ049 ADYJ051 ADYJ054 ADYJ056 ADYJ061 ADYJ062 ADYJ064 through ADYJ067 ADYJ069 through ADYJ073 ADYJ075 through ADYJ082 AEYJ001 through AEYJ003 AEYJ005 AEYJ006 AEYJ008 through AEYJ010 AEYP003 through AEYP011 AEYP024 through AEYP026 AEYP028 through AEYP030 AFYP005 through AFYP008 AFYP012 AFYP014 through AFYP018 AFYP021 AFYP022 AGYP010 AGYP012 AGYP016 AGYP018 through AGYP020 AGYP022 AGYP025 AGYP027 AGYP028 AGYP030 AGYP031 AGYP033 through AGYP041 AGYP045 AGYP046 AGYP068 AGYP069 AGYP073 AGYP086 AGYP088 AGYP089 AGYP092 AJYP025 AJYP027 AKYP001 AKYP007 AKYP009. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
There have been approximately 2% of the x-ray generators manufactured between February 2012 and November 2012 returned due to errors. These generators showed the type of defect that indicates poor contact of HT transformer power pin. The defect is a bent connector associated with the generator.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Trophy Sas
- Distribuidor
- USA Nationwide Distribution
- Lote
- ABYJ011 ABYJ012 ABYJ020 through ABYJ026 ACJY522 ACYJ004 through ACYJ007 ACYJ009 through ACYJ012 ACYJ014 ACYJ015 ACYJ017 through ACYJ023 ACYJ500 through ACYJ509 ACYJ517 through ACYJ521 ACYJ523 through ACYJ531 ACYJ533 through ACYJ536 ADYJ001 through ADYJ005 ADYJ007 ADYJ008 ADYJ011 through ADYJ013 ADYJ015 through ADYJ018 ADYJ020 through ADYJ032 ADYJ034 through ADYJ037 ADYJ039 through ADYJ049 ADYJ051 ADYJ054 ADYJ056 ADYJ061 ADYJ062 ADYJ064 through ADYJ067 ADYJ069 through ADYJ073 ADYJ075 through ADYJ082 AEYJ001 through AEYJ003 AEYJ005 AEYJ006 AEYJ008 through AEYJ010 AEYP003 through AEYP011 AEYP024 through AEYP026 AEYP028 through AEYP030 AFYP005 through AFYP008 AFYP012 AFYP014 through AFYP018 AFYP021 AFYP022 AGYP010 AGYP012 AGYP016 AGYP018 through AGYP020 AGYP022 AGYP025 AGYP027 AGYP028 AGYP030 AGYP031 AGYP033 through AGYP041 AGYP045 AGYP046 AGYP068 AGYP069 AGYP073 AGYP086 AGYP088 AGYP089 AGYP092 AJYP025 AJYP027 AKYP001 AKYP007 AKYP009
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- France
Retiros relacionados
- FDACareStream Dental Kodak 2200 Intraoral X-ray system Product Usage: The expected usage (intended use) for all8 de mayo de 2013
- FDACareStream Dental Kodak 2100 Intraoral X-ray system Product Usage: The expected usage (intended use) for all8 de mayo de 2013
- FDACareStream Dental CS 2200 X-ray system Product Usage: The expected usage (intended use) for all of these8 de mayo de 2013
- FDABUFFALO TRACE HOT BARBEQUE SAUCE Net Wt. 16 fl. oz./532 ml PRODUCED BY: APPLECREEK ORCHARDS LLC P.O. Box 83838 de mayo de 2013
- FDABUFFALO TRACE ORIGINAL FLAVOR BARBEQUE SAUCE Net Wt. 16 fl. oz./532 ml PRODUCED BY: APPLECREEK ORCHARDS LLC8 de mayo de 2013
- FDABUFFALO TRACE MARINADE FOR MEAT Net Wt. 12 fl. oz. PRODUCED BY: APPLECREEK ORCHARDS LLC P.O. Box 83838 de mayo de 2013
Preguntas sobre este retiro
¿CareStream Dental CS 2100 X-ray system Product Usage: The expected usage (intended use) for all of these products is for dental radiographic examination and diagnosis of the teeth, jaw and oral structures. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0999-2013 el 8 de mayo de 2013. Peligro reportado: There have been approximately 2% of the x-ray generators manufactured between February 2012 and November 2012 returned due to errors. These generators showed the type of defect that indicates poor contact of HT transformer power pin. The defect is a bent connector associated with the generator.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter