AirLife Adult Manual Resuscitator, 40" (1.0 m) Oxygen Reservoir Tubing, Adult Mask, CO2 Detector, REF 2K8005C2

Publicado el 6 de marzo de 2024 · Inicio del retiro: 10 de enero de 2024

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Vyaire Medical
Distribuidor
Worldwide distribution. US nationwide, Brazil, Canada, Switzerland, France, UK, Mexico, Netherlands, Saudi Arabia, Sweden, and South Africa
Lote
UDI/DI Case: 50190752114164, Each: 10190752114166; All manufacturing dates of 2017 and prior, and any products without a manufacturing date.
Clasificación de riesgo
Class I