Siemens brand ACUSON SC2000 Ultrasound System, Model 10433816, hardware that contains version 4 of the MPI board 10434200; The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transespohageal, Adult Cephalic, Peripheral Vessel, Abdominal, Abdominal Intraoperative, Intraoperative Neurological, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
Publicado el 17 de abril de 2013 · Inicio del retiro: 28 de marzo de 2013
Resumen
FDA registra el retiro Z-1081-2013 de dispositivos médicos de Clase II, publicado el 17 de abril de 2013 e iniciado el 28 de marzo de 2013. Empresa responsable: Siemens Medical Solutions USA, Inc.; Estado: Finalizado; Cantidad reportada: 120 units; Lote: All units of this model. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
When using the V5M transducer, the display of the temperature on the ACUSON SC2000 system can be lower than the actual temperature of the V5M transducer. This occurs only if the system hardware contains revision 4 of the MPI Board. There is a potential to cause esophageal burns in patients.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Siemens Medical Solutions USA, Inc.
- Distribuidor
- Worldwide Distribution - USA and countries of Australia, Austria, Bosnia Herzegovina, Belgium, Finland, France, Germany, India, Italy, Japan, Kuwait, Malaysia, Netherlands, Norway, P.R. China, Portugal, Rep.of Korea, Romania, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan and United Kingdom.
- Lote
- All units of this model.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAACUSON Redwood 1.0 and 2.0 Diagnostic Ultrasound Systems, REF: 11503314, with software2 de octubre de 2024
- FDAACUSON Juniper 1.0, 1.5, 2.0, 2.5 Diagnostic Ultrasound Systems, REF: 11335791, with software. ACUSON Juniper2 de octubre de 2024
- FDAACUSON Maple 1.0 Diagnostic Ultrasound System, REF: 11711750, with software. ACUSON Maple Select 1.02 de octubre de 2024
- FDAACUSON Redwood 1.0 Diagnostic Ultrasound System, REF: 11503314, running software version VA10A to VA10G1 de mayo de 2024
- FDACO2 Incubators (General Purpose Incubator), model numbers 370, 371, 380, 381, 3307, 3308, 3310, 3311, 3540,17 de abril de 2013
- FDAThe LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators. Product part's numbers: 3200731-xxx, 3202177-xxx,17 de abril de 2013
Preguntas sobre este retiro
¿Siemens brand ACUSON SC2000 Ultrasound System, Model 10433816, hardware that contains version 4 of the MPI board 10434200; The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transespohageal, Adult Cephalic, Peripheral Vessel, Abdominal, Abdominal Intraoperative, Intraoperative Neurological, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1081-2013 el 17 de abril de 2013. Peligro reportado: When using the V5M transducer, the display of the temperature on the ACUSON SC2000 system can be lower than the actual temperature of the V5M transducer. This occurs only if the system hardware contains revision 4 of the MPI Board. There is a potential to cause esophageal burns in patients.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter