Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1207-2015

NAMIC Convenience Kit, RADIOLOGY KIT, UPN H7496002072111, REF/Catalog No. 600207211, STERILE, Rx ONLY --- For single use only. Device Listing Number D136683. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.

Publicado el 4 de marzo de 2015 · Inicio del retiro: 9 de enero de 2015

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Navilyst Medical, Inc
Distribuidor
Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
Lote
Lots 4244975, 4251615, 4269276, 4292593, 4310081, 4344776, 4374583, 4384503, 4413658, 4447492, 4505410, 4535348, 4569287, 4572788, 4604137, 4610092, 4612832, 4630601, 4630602, 4638044, 4681013, 4690175, 4705330, 4717942, 4731420, 4772544, 4812433.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States