FDAFDA · Alimentos y medicamentos·N.º oficial Z-1227-2024
VITROS XT 3400 Chemistry System, Catalog No. 6844458
Publicado el 13 de marzo de 2024 · Inicio del retiro: 26 de enero de 2024
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Ortho-Clinical Diagnostics, Inc.
- Distribuidor
- Worldwide - US Nationwide distribution in the states of AZ, CA, CT, FL, IA, ID, IL, IN, KY, LA, MO, NE, NY, OH, OR, SC, SD, TX, WA, WI, WV and the countries of Australia, Brazil, Canada, China, Colombia, India, Mexico, Singapore, United Kingdom, Italy, Spain, Russia, France, and Belgium.
- Lote
- VITROS System Software Version 3.8.1 UDI-DI 10758750031986 Serial Numbers US: J34500180 J34500289 J34500422 J34500473 J34500622 J34500861 J34500887 OUS: J34500515 J34500609 J34500802 J34500940
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States