MRI system: Vantage Elan, Model: MRT-2020
Publicado el 13 de marzo de 2024 · Inicio del retiro: 29 de enero de 2024
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Canon Medical System, USA, INC.
- Distribuidor
- Worldwide - US Nationwide distribution including in the states of MD, IL, WV, CA, AZ, PR, CO, GA, PA, WA, UT, OK, KY, IA, IN, TN, CT, WI, OH, AL, MT, NM, TX, MO, MN, OR, ID, WY, NV, AR, FL, SD, MI, NY, NJ, KS, LA, NH, VT, MS, SC, ME and the country of DO.
- Lote
- UDI-DI: 04987670101632. Serial Number: S1F2052753
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States