Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1274-2013

Zilver® PTX® Drug-Eluting Peripheral Stent - 6 mm X 80mm; 125 cm, Rx, Sterile Product Usage: The Zilver PTX Drug -Eluting Peripheral Stent is a self expanding, small metal, mesh tube (stent) with the outer surface coated with the drug Paclitaxel that can be Implanted in an artery in the thigh (femoropopliteal artery). The Paclitaxel coating helps Preven the artery from narrowing again (restenosis). The Zilver; PTX stent acts as a scaffold to:  Hold open a narrowed artery in the thigh (femoropopliteal artery) caused by PAD .  improve blood flow to the extremity.

Publicado el 22 de mayo de 2013 · Inicio del retiro: 18 de abril de 2013

Resumen

FDA registra el retiro Z-1274-2013 de dispositivos médicos de Clase I, publicado el 22 de mayo de 2013 e iniciado el 18 de abril de 2013. Empresa responsable: Cook, Inc.; Estado: Finalizado; Cantidad reportada: 1,057 units in USA; Lote: REF ZIV6-35-125-6-80-PTX ; REFG24890 UDI 10827002248905 All lots distributed prior to 4/16/2013. La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Cook Medical has received a small number of complaints related to the delivery system for the Zilver PTX Drug-Eluting Peripheral Stent involving fractures of the delivery system inner catheter after stent deployment, and separation of the inner catheter tip section (13 complaints with an occurrence rate of 0.043%) Two adverse events, including one death, occurred in cases where a tip separation

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Cook, Inc.
Distribuidor
Worldwide Distribution - USA Nationwide; in the countries of Japan, Europe, Brazil, and Australia
Lote
REF ZIV6-35-125-6-80-PTX ; REFG24890 UDI 10827002248905 All lots distributed prior to 4/16/2013
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Zilver® PTX® Drug-Eluting Peripheral Stent - 6 mm X 80mm; 125 cm, Rx, Sterile Product Usage: The Zilver PTX Drug -Eluting Peripheral Stent is a self expanding, small metal, mesh tube (stent) with the outer surface coated with the drug Paclitaxel that can be Implanted in an artery in the thigh (femoropopliteal artery). The Paclitaxel coating helps Preven the artery from narrowing again (restenosis). The Zilver; PTX stent acts as a scaffold to:  Hold open a narrowed artery in the thigh (femoropopliteal artery) caused by PAD .  improve blood flow to the extremity. fue retirado del mercado?

Sí. FDA publicó el retiro Z-1274-2013 el 22 de mayo de 2013. Peligro reportado: Cook Medical has received a small number of complaints related to the delivery system for the Zilver PTX Drug-Eluting Peripheral Stent involving fractures of the delivery system inner catheter after stent deployment, and separation of the inner catheter tip section (13 complaints with an occurrence rate of 0.043%) Two adverse events, including one death, occurred in cases where a tip separation

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter