Brand Name: Molift Product Name: Molift 4-point sling bars (Accessory to mobile hoists) Model/Catalog Number: 1430116,1430117,1430115 Product Description: The mobile hoist is an assistive device intended for alleviation of, or compensation for, a functional, impairment due to an injury or disability. The device is designed for an individual lacking the ability to stand up and transfer themselves to / from a bed, the floor, a wheelchair, a chair, a toilet or similar due to reduced mobility or physical strength.
Publicado el 11 de febrero de 2026 · Inicio del retiro: 17 de noviembre de 2025
Resumen
FDA registra el retiro Z-1311-2026 de dispositivos médicos de Clase II, publicado el 11 de febrero de 2026 e iniciado el 17 de noviembre de 2025. Empresa responsable: ETAC A/S; Estado: En curso; Cantidad reportada: 25 units; Lote: (1) Model/Catalog Number: 1430116; UDI-DI: TBD; (2) Model/Catalog Number: 1430117; UDI-DI: TBD; (3) Model/Catalog Number: 1430115; UDI-DI: TBD. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.
Qué hacer
Voluntary: Firm initiated · Telephone
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- ETAC A/S
- Distribuidor
- Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland, France, Germany, Israel, Italy, Japan, New Zealand, Norway, Poland, Qatar Spain, Sweden, Switzerland, UK.
- Lote
- (1) Model/Catalog Number: 1430116; UDI-DI: TBD; (2) Model/Catalog Number: 1430117; UDI-DI: TBD; (3) Model/Catalog Number: 1430115; UDI-DI: TBD;
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- Denmark
Retiros relacionados
- FDAPro Numb Tattoo Numbing Spray, 5% Lidocaine, [1 FL OZ/28.4 ML or 4 oz/120 mL] per bottle, Distributed by: Pro18 de febrero de 2026
- FDAPro Numb Tattoo Numbing Spray, For Sensitive Skin, 4% Lidocaine, [1FL OZ/28.4 ML or 4 oz/120 mL] per bottle,18 de febrero de 2026
- FDADoxazosin Tablets, USP, 4 mg, 1000 tablets per bottle, Rx only, distributed by Unichem Pharmaceuticals (USA),18 de febrero de 2026
- FDAProduct Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable11 de febrero de 2026
- FDAProduct Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description:11 de febrero de 2026
- FDAProduct Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description:11 de febrero de 2026
Preguntas sobre este retiro
¿Brand Name: Molift Product Name: Molift 4-point sling bars (Accessory to mobile hoists) Model/Catalog Number: 1430116,1430117,1430115 Product Description: The mobile hoist is an assistive device intended for alleviation of, or compensation for, a functional, impairment due to an injury or disability. The device is designed for an individual lacking the ability to stand up and transfer themselves to / from a bed, the floor, a wheelchair, a chair, a toilet or similar due to reduced mobility or physical strength. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1311-2026 el 11 de febrero de 2026. Peligro reportado: A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Telephone