FDAFDA · Alimentos y medicamentos·N.º oficial Z-1435-2017
MEVATRON M 7400/7440(MEX) Accelerator, Linear, Medical The intended use of the SIEMENS branded PRIMUS family of linear accelerator systems is to deliver x-ray radiation for therapeutic treatment of cancer
Publicado el 15 de marzo de 2017 · Inicio del retiro: 25 de enero de 2017
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Siemens became aware of a defective weld seam at overhead suspensions. Due to this defect it is possible for the weld seam to break and to cause the overhead suspension to fall down. This can lead to a severe injury to a patient or any other person.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Siemens Medical Solutions USA, Inc
- Distribuidor
- Nationwide Distribution
- Lote
- Device Model Number 8319758 Serial numbers: 3009 2958 2863 2972 2646 3079 2516 2962 3055 1452 2702 2125 3066 2689 2228 2564 2507 1904 3075 2375 3013 3068 2596 3025 3049 2965 2045 3084 2855 2906 2292 2384 2067 2818 3089
- Clasificación de riesgo
- Class II