adult defibrillation, cardioversion, pacing, monitoring Disposable Electrodes, Model P-211-M1 and P-214-M1. Product Usage: Used by trained professionals in hospitals, doctors offices, and Emergency Medical Services for adult low-energy defibrillation, transcutaneous pacing, cardioversion and monitoring.

Publicado el 5 de junio de 2013 · Inicio del retiro: 30 de abril de 2013

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

TZ Medical Inc is recalling its P-211-M1 and P-214-M1 Medtronic / Physio Control Compatible transparent defibrillation electrodes because it may contain a defective white wire on the connector assembly. If a failure were to occur, the end user would be able to identify the failure by inspecting the white wire 1/4 inch from the Gray Medtronic / Physio adapter. The consignees will notice discolor

Qué hacer

Voluntary: Firm initiated · Telephone

Detalles del retiro

Fabricante
TZ Medical Inc.
Distribuidor
Worldwide distribution - USA (nationwide) and Chile.
Lote
Model P-211-M1 Lot numbers: Y081712-17; Y081712-18; Y081712-19; Y081712-20; Y112612-08; Y112612-09; Y112612-10; Y112612-11; Y112612-12; Y112612-13; and Y020113-07. Model P-214-M1 Lot numbers: Y081712-04; Y112612-24; Y112612-23; and Y031913-10.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States