Part 510031 Disposable Single Infusion Pump, quantity Includes... TOTAL VOLUME: 150mL, 2 mL/hr continuous Delivery Time : App. 72 hrs when filled to capacity Sterile EO... found in the following GoPump Kits: GoPump Kit 510042, w/ Epidural Catheter GoPump Kit 510042 - BP, w/ Epidural Catheter, BIOPATCH GoPump Kit 510110, w/2.5" Fenestrated Catheter GoPumpKit 510110-BP, w/2.5" Fenestrated Catheter,BioPatch GoPumpKit 510350 w/ Epidural Catheter GoPump Kit 510350-BP w/ Epidural Catheter, BIOPATCH GoPumpKit 510556-BP , 2.5 catheter, BIOPATCH Symbios GOPump Rapid Recovery System is a local pain mangagement system that consits of a small balloon that is inflated with a local anesthetic medication.
Publicado el 12 de junio de 2013 · Inicio del retiro: 10 de mayo de 2013
Resumen
FDA registra el retiro Z-1452-2013 de dispositivos médicos de Clase I, publicado el 12 de junio de 2013 e iniciado el 10 de mayo de 2013. Empresa responsable: Symbios Medical Products, LLC; Estado: Finalizado; Cantidad reportada: 33, 543 in all sizes/ kits; Lote: Pump 510031, Lot 11-l 00106 to Lot 12-101133 Kit 510042, lots 11-100251 to 13-100128 Kit 510042-BP , lot 12-101300 Kit 510110, lot 11-100664 Kit 510110-BP, Lots 11-100838 to 12-101560 Kit 510350, Lots 11-100353 to 12-100233 Kit 510350-BP, Lots12-100429 to 13-100047 Kit 510556-BP Lots 11-100779 to 12-101552. La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Symbios Medical Products, LLC
- Distribuidor
- US Nationwide distribution.
- Lote
- Pump 510031, Lot 11-l 00106 to Lot 12-101133 Kit 510042, lots 11-100251 to 13-100128 Kit 510042-BP , lot 12-101300 Kit 510110, lot 11-100664 Kit 510110-BP, Lots 11-100838 to 12-101560 Kit 510350, Lots 11-100353 to 12-100233 Kit 510350-BP, Lots12-100429 to 13-100047 Kit 510556-BP Lots 11-100779 to 12-101552
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAPart 510449 Disposable Infusion Pump, ...Elastomeric Pump 600 mL( 8mL/hr ),... found in the following kits12 de junio de 2013
- FDAPart 510636 Disposable Infusion Pump, ...Elastomeric Pump 600 mL( 300 ml perside), 2 to 10 ml/hr. Delivery12 de junio de 2013
- FDAPart 510180 Disposable Single Infusion Pump, ... Elastomeric Pump 300 mL, 2mL/hr,... found in : GoPump Kit12 de junio de 2013
- FDAPart 510490 Disposable Infusion Pump, ...Elastomeric Pump 600 mL( 300 ml per side)( 6mL/hr ),... found in the12 de junio de 2013
- FDAComfort brand Ear-Loop Face Mask, Item #: 3100, Model Numbers: # 3100B, # 3100P, # 3100G, # 3100Y, # 3100W;12 de junio de 2013
- FDAPart 510076 Disposable Dual Infusion Pump, ...Elastomeric Dual Pump 300 mL(150 ml/side) 2mL/hr/side,...12 de junio de 2013
Preguntas sobre este retiro
¿Part 510031 Disposable Single Infusion Pump, quantity Includes... TOTAL VOLUME: 150mL, 2 mL/hr continuous Delivery Time : App. 72 hrs when filled to capacity Sterile EO... found in the following GoPump Kits: GoPump Kit 510042, w/ Epidural Catheter GoPump Kit 510042 - BP, w/ Epidural Catheter, BIOPATCH GoPump Kit 510110, w/2.5" Fenestrated Catheter GoPumpKit 510110-BP, w/2.5" Fenestrated Catheter,BioPatch GoPumpKit 510350 w/ Epidural Catheter GoPump Kit 510350-BP w/ Epidural Catheter, BIOPATCH GoPumpKit 510556-BP , 2.5 catheter, BIOPATCH Symbios GOPump Rapid Recovery System is a local pain mangagement system that consits of a small balloon that is inflated with a local anesthetic medication. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1452-2013 el 12 de junio de 2013. Peligro reportado: Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter