ExacTrac 6.0 is a patient positioning and monitoring system. Model/catalogue numbers: 20833B EXACTRAC 6.0 IR POSITIONING SOFTWARE 49936 ET SOFTWARE UPDATE 6.0.X TO 6.0.3 49926B ET UPGRADE SOFTWARE 3.X TO 6.0 (IR+XR) 49934A EXACTRAC UPGRADE 5.5 TO 6.0 BASIC 49927B ET UPGRADE SOFTWARE 4.X TO 6.0 (IR+XR) 49928B ET UPGRADE SOFTWARE 5.X TO 6.0 (IR+XR) 49933B ET UPGRADE TRUEBEAM 5.5 TO 6.0 LIMITED 49973B ET DATA PREP / REVIEW SYSTEM 49998B ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT 49996B ET UPGRADE SOFTWARE X.X TO 6.0 SW ONLY 49997B ET UPGRADE SOFTWARE 5.5 TO 6.0 LIMITED
Publicado el 7 de mayo de 2014 · Inicio del retiro: 24 de febrero de 2014
Resumen
FDA registra el retiro Z-1513-2014 de dispositivos médicos de Clase II, publicado el 7 de mayo de 2014 e iniciado el 24 de febrero de 2014. Empresa responsable: Brainlab AG; Estado: Finalizado; Cantidad reportada: 41 systems (13 systems in US; 28 systems to foreign countries); Lote: ExacTrac version v.6.0.3. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
When using multiple radiation treatment targets within a single plan, the ExacTrac v.6.0.3 might position the patient to an unintended radiation treatment point. As a result, there is a possibility that the radiation treatment dose may be delivered to the unintended target position.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Brainlab AG
- Distribuidor
- Worldwide Distribution -- USA, including the states of CA, IN, LA, MA, OH, TX, and WI, and the countries of Australia, China, France, Germany, Japan, Netherlands, and the United Kingdom.
- Lote
- ExacTrac version v.6.0.3.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- Germany
Retiros relacionados
- FDAUltrasound Navigation Software 1.0 and Ultrasound Integration Software for Cranial/ENT Navigation 3.0.x and14 de octubre de 2020
- FDASpine & Trauma 3D Navigation 1.0 (subpart of the system Navigation Software Spine & Trauma 3D, Version 3.0 )24 de abril de 2019
- FDAExacTrac versions 6.5 through 6.5 intended to be used to place patients at an accurately defined point within21 de marzo de 2018
- FDAAIRO Mobile CT System, Model Number: MobiCT-32 Product Usage: The AIRO is intended to be used for X-ray14 de febrero de 2018
- FDAGemStar Docking Station, used with GemStar infusion pump, in combination with other GemStar accessories such7 de mayo de 2014
- FDAHexaPOD evo RT Couchtop. To support and aid in positioning a patient during radiation therapy.7 de mayo de 2014
Preguntas sobre este retiro
¿ExacTrac 6.0 is a patient positioning and monitoring system. Model/catalogue numbers: 20833B EXACTRAC 6.0 IR POSITIONING SOFTWARE 49936 ET SOFTWARE UPDATE 6.0.X TO 6.0.3 49926B ET UPGRADE SOFTWARE 3.X TO 6.0 (IR+XR) 49934A EXACTRAC UPGRADE 5.5 TO 6.0 BASIC 49927B ET UPGRADE SOFTWARE 4.X TO 6.0 (IR+XR) 49928B ET UPGRADE SOFTWARE 5.X TO 6.0 (IR+XR) 49933B ET UPGRADE TRUEBEAM 5.5 TO 6.0 LIMITED 49973B ET DATA PREP / REVIEW SYSTEM 49998B ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT 49996B ET UPGRADE SOFTWARE X.X TO 6.0 SW ONLY 49997B ET UPGRADE SOFTWARE 5.5 TO 6.0 LIMITED fue retirado del mercado?
Sí. FDA publicó el retiro Z-1513-2014 el 7 de mayo de 2014. Peligro reportado: When using multiple radiation treatment targets within a single plan, the ExacTrac v.6.0.3 might position the patient to an unintended radiation treatment point. As a result, there is a possibility that the radiation treatment dose may be delivered to the unintended target position.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter