The Osteo-Clage System consists of stainless steel cable/sleeves and compression plates. . 2.0 mm x 850mm SS Cable/Sleeve; Part Number 01-0020-S 2.0 mm SS Sleeve Only; Part Number 01-0022-S 8, 7 hole, SS compression plate; Part Number 02-2007-S 10, 9 hole, SS compression plate; Part Number 02-2009-S Product Usage: The Osteo-Clage Stainless Steel Bone Plate is an implantable straight, rigid stainless steel plate. It is used in conjunction with 4.5mm Cortical bone screws to provide compression across bone fractures. The Osteo-Clage Cerclage Wire Crimp sleeve is used in conjunction with the bone plate in cerclage fixation procedures
Publicado el 7 de mayo de 2014 · Inicio del retiro: 26 de marzo de 2014
Resumen
FDA registra el retiro Z-1515-2014 de dispositivos médicos de Clase II, publicado el 7 de mayo de 2014 e iniciado el 26 de marzo de 2014. Empresa responsable: Acumed LLC; Estado: Finalizado; Cantidad reportada: 141 units in the US; 294 units outside the US.; Lote: 2.0mm x 80mm SS cable/sleeve: Lot # 222078, 232175, 245032, 254368, 261903, 270659, 284155, 297281. 2.0mm SS sleeve only: Lot # 247553, 267469, 280871, 313261. 8, 7 hole SS compression plate: Lot # 300381. 10, 9 hole SS compression plate: Lot # 299641. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Manufacturing of these devices with a grade of stainless steel that is not within specifications.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Acumed LLC
- Distribuidor
- US Nationwide Distribution - in the states of CA, CO, FL, GA, LA, MI, MN, MO, OH, OR, PA, PR, TX, UT, VA, WA and WY. Product was distributed outside the US to: Australia, China, Figi, France, Great Britain, Italy, Japan, Korea, Maylasia, Sweden, South Africa, Spain.
- Lote
- 2.0mm x 80mm SS cable/sleeve: Lot # 222078, 232175, 245032, 254368, 261903, 270659, 284155, 297281. 2.0mm SS sleeve only: Lot # 247553, 267469, 280871, 313261. 8, 7 hole SS compression plate: Lot # 300381. 10, 9 hole SS compression plate: Lot # 299641.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAAcumed Hexalob Screw various sizes: REF Numbers: : 3070-27008 2.7mm x 8mm NL Hexalobe Screw; 3070-27009 2.7mm1 de julio de 2026
- FDAAcu-Loc and Acu-Loc 2 bone plates, screws and accessories for the following REF: 70-0045 Acu-Loc VDU Plate6 de marzo de 2024
- FDARibLoc U Plus 90 Instrument PRIMARY GUIDE ASSY, LOW PROFILE REF RBL2320 The RBL2320, Low Profile Primary Guide3 de enero de 2024
- FDAThe Acumed Slide-Loc Anatomic Radial Head System is intended for use by surgeons with orthopedic training. The7 de junio de 2017
- FDAHexaPOD evo RT Couchtop. To support and aid in positioning a patient during radiation therapy.7 de mayo de 2014
- FDAThe Acumed Tension Band Pin is a fixation pin which is .0625" in diameter and is available in lengths between7 de mayo de 2014
Preguntas sobre este retiro
¿The Osteo-Clage System consists of stainless steel cable/sleeves and compression plates. . 2.0 mm x 850mm SS Cable/Sleeve; Part Number 01-0020-S 2.0 mm SS Sleeve Only; Part Number 01-0022-S 8, 7 hole, SS compression plate; Part Number 02-2007-S 10, 9 hole, SS compression plate; Part Number 02-2009-S Product Usage: The Osteo-Clage Stainless Steel Bone Plate is an implantable straight, rigid stainless steel plate. It is used in conjunction with 4.5mm Cortical bone screws to provide compression across bone fractures. The Osteo-Clage Cerclage Wire Crimp sleeve is used in conjunction with the bone plate in cerclage fixation procedures fue retirado del mercado?
Sí. FDA publicó el retiro Z-1515-2014 el 7 de mayo de 2014. Peligro reportado: Manufacturing of these devices with a grade of stainless steel that is not within specifications.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter