Giraffe® Infant Warmers integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable to thermo-regulate based on their own physiology with integrated monitoring feature used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by a SpO2 sensor).
Publicado el 14 de mayo de 2014 · Inicio del retiro: 2 de abril de 2014
Resumen
FDA registra el retiro Z-1547-2014 de dispositivos médicos de Clase II, publicado el 14 de mayo de 2014 e iniciado el 2 de abril de 2014. Empresa responsable: GE Healthcare; Estado: Finalizado; Cantidad reportada: 50 units; Lote: HDJM50062 HDJM50093 HDJM50470 HDJM50654 HDJM51812 HDJS55099 HDJS53396 HDJS53397 HDJS53398 HDJT50393 HDJS55036 HDJS55037 HDJS55038 HDJT50318 HDJS52165 HDJS52166 HDJS52119 HDJS52120 HDJR57563 HDJR57564 HDJR57565 HDJR57566 HDJR57567 HDJR57568 HDJR57569 HDJS50733 HDJS52092 HDJS52093 HDJS51002 HDJS51003 HDJS51004 HDJS51005 HDJS51006 HDJP50601 HDJP50606 HDJN50376 HDJS50790 HDJS50791 HDJS50792 HDJS50159 HDJS50160 HDJS51426 HDJS51427 HDJS50734 HDJS50735 HDJS52118 HDJT50534 HDJT50039 HDJR58198 HDJS50316. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- GE Healthcare
- Distribuidor
- Worldwide Distribution.
- Lote
- HDJM50062 HDJM50093 HDJM50470 HDJM50654 HDJM51812 HDJS55099 HDJS53396 HDJS53397 HDJS53398 HDJT50393 HDJS55036 HDJS55037 HDJS55038 HDJT50318 HDJS52165 HDJS52166 HDJS52119 HDJS52120 HDJR57563 HDJR57564 HDJR57565 HDJR57566 HDJR57567 HDJR57568 HDJR57569 HDJS50733 HDJS52092 HDJS52093 HDJS51002 HDJS51003 HDJS51004 HDJS51005 HDJS51006 HDJP50601 HDJP50606 HDJN50376 HDJS50790 HDJS50791 HDJS50792 HDJS50159 HDJS50160 HDJS51426 HDJS51427 HDJS50734 HDJS50735 HDJS52118 HDJT50534 HDJT50039 HDJR58198 HDJS50316
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDACentricity Enterprise Archive V2, V3, and V4, software product for receiving, archiving, and sending23 de septiembre de 2026
- FDAWorkflow Core Services V8, intended to provide set of services for storage and retrieval of patient data,23 de septiembre de 2026
- FDACentricity PACS V2, V3, V4, V6, V7, software product is intended for the storage, reading, diagnostic review,23 de septiembre de 2026
- FDACore Services Hub V8, a set of backend services including Cross Enterprise (XE) and Workflow Core (WFC)23 de septiembre de 2026
- FDAAlere INRatio2 PT/INR Professional Test Strips, Model Number: 99008G2. The Alere INRatio2 PT/INR Monitoring14 de mayo de 2014
- FDAPanda® iRes Infant Warmers integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable14 de mayo de 2014
Preguntas sobre este retiro
¿Giraffe® Infant Warmers integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable to thermo-regulate based on their own physiology with integrated monitoring feature used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by a SpO2 sensor). fue retirado del mercado?
Sí. FDA publicó el retiro Z-1547-2014 el 14 de mayo de 2014. Peligro reportado: Medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter