Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1575-2013

Medtronic INDURA 1P Intrathecal Catheter, model 8709SC. Contents: 81.4-cm catheter with length markers and guide wire, 15 T-gauge introducer needle, Sutureless pump connector with attached 7.6-cm catheter and Connector pin, Transparent Strain-relief sleeves, and anchors. Contents of inner package are STERILE. The implantable Medtronic INDURA 1P Model 8709SC Intrathecal Catheter is part of an infusion system that stores and delivers parenteral drugs to the intrathecal space. The implanted infusion system components consist of a Medtronic pump and a Model 8709SC catheter. The catheter connects to the pump at the catheter port.

Publicado el 3 de julio de 2013 · Inicio del retiro: 3 de junio de 2013

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The Sutureless Connector (SC) Intrathecal Catheter connector has been redesigned to reduce the potential for occlusion at the catheter to pump interface. Medtronic is removing the unused products from the market that were manufactured with the previous design, and recommend the previous design no longer be used due to greater potential for misalignment and subsequent occlusion.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Medtronic Neuromodulation
Distribuidor
Worldwide distribution: US (nationwide) and countries of: Aruba, Australia, Austria, Belgium, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, Georgia, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Jordan, Korea, Kuwait, Lebanon, Lithuania, Luxembourg, Malta, Martinique, Morocco, Netherlands, Netherlands Antilles, New Zealand, Norway, Panama, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, Uruguay, and Venezuela.
Lote
Product having a Use By Date prior to 25 Aug 2014
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta