HCS - Headless Compression Screw System, Headless Compression Screw, Cannulated; Article Numbers: 31207-10, 31207-12, 31207-14, 31207-16, 31207-18, 31207-20, 31207-24, 31207-26, 31207-28, 31207-30, 31257-12, 31257-14, 31257-16, 31257-18, 31257-20, 31257-22, 31257-24, 31257-28, 31257-30, 31307-18, 31307-32, 31307-34, 31307-36, 31307-38, 31358-30, 31437-30, 31437-38, 31437-40, 31437-50, 31757-100, 31757-120, 31757-45, 31757-80, 31757-85, 31757-90.
Publicado el 1 de abril de 2026 · Inicio del retiro: 23 de enero de 2026
Resumen
FDA registra el retiro Z-1606-2026 de dispositivos médicos de Clase II, publicado el 1 de abril de 2026 e iniciado el 23 de enero de 2026. Empresa responsable: I.T.S. GmbH; Estado: En curso; Cantidad reportada: 623 units; Lote: Article Number (Lot Numbers): 31207-10 (679/112002), 31207-12 (679/092133, 679/102343, 679/112006), 31207-14 (679/102050), 31207-16 (679/102345, 679/102049), 31207-18 (679/0323110, 679/102355, 679/112001), 31207-20 (679/0323111, 679/102051, 679/102356), 31207-24 (679/102028), 31207-26 (679/112008), 31207-28 (679/112009), 31207-30 (679/0323140, 86/483377), 31257-12 (679/081904), 31257-14 (679/042323, 679/081905), 31257-16 (679/042324), 31257-18 (679/112308, 679/022128), 31257-20 (679/022123, 679/042326, 679/112321), 31257-22 (679/022124, 679/042327, 679/112325), 31257-24 (679/022125, 679/112326), 31257-28 (679/022130, 679/052306), 31257-30 (679/022131), 31307-18 (679/061907), 31307-32 (679/062285), 31307-34 (679/062286), 31307-36 (679/062287), 31307-38 (679/032118), 31358-30 (26/F67790), 31437-30 (679/071916), 31437-38 (679/071918), 31437-40 (679/032406, 679/071903), 31437-50 (679/071920), 31757-100 (35/51892), 31757-120 (86/320792), 31757-45 (86/320793), 31757-80 (26/F55872), 31757-85 (26/F55873), 31757-90 (26/F55874). La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Qué hacer
Voluntary: Firm initiated · E-Mail
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- I.T.S. GmbH
- Distribuidor
- US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
- Lote
- Article Number (Lot Numbers): 31207-10 (679/112002), 31207-12 (679/092133, 679/102343, 679/112006), 31207-14 (679/102050), 31207-16 (679/102345, 679/102049), 31207-18 (679/0323110, 679/102355, 679/112001), 31207-20 (679/0323111, 679/102051, 679/102356), 31207-24 (679/102028), 31207-26 (679/112008), 31207-28 (679/112009), 31207-30 (679/0323140, 86/483377), 31257-12 (679/081904), 31257-14 (679/042323, 679/081905), 31257-16 (679/042324), 31257-18 (679/112308, 679/022128), 31257-20 (679/022123, 679/042326, 679/112321), 31257-22 (679/022124, 679/042327, 679/112325), 31257-24 (679/022125, 679/112326), 31257-28 (679/022130, 679/052306), 31257-30 (679/022131), 31307-18 (679/061907), 31307-32 (679/062285), 31307-34 (679/062286), 31307-36 (679/062287), 31307-38 (679/032118), 31358-30 (26/F67790), 31437-30 (679/071916), 31437-38 (679/071918), 31437-40 (679/032406, 679/071903), 31437-50 (679/071920), 31757-100 (35/51892), 31757-120 (86/320792), 31757-45 (86/320793), 31757-80 (26/F55872), 31757-85 (26/F55873), 31757-90 (26/F55874).
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- Austria
Retiros relacionados
- FDAOlecranonplate with Angular Stability with the below descriptions and corresponding article numbers. 1.1 de abril de 2026
- FDAFLS - Foot Locking Plates System and HOL - Hallux Osteotomy Locking Plate with the below descriptions and1 de abril de 2026
- FDAPhilips Respironics Trilogy EV300, Software Version 1.05.15.00. Continuous home-use ventilator device.1 de abril de 2026
- FDAPhilips Respironics Trilogy Evo Universal, Software Version 1.05.15.00. Continuous home-use ventilator device.1 de abril de 2026
- FDAFlexible Cryoprobe (OD 1.1mm, L1.15mm) w/ oversheath (OD2.6mm, L817mm) REF: 20402-401 STERILE EO. For surgical1 de abril de 2026
- FDADrill Plate, Ortho Dev. Trimax - Pegged, PREF Size 1 -5 REF:402566 component Lantern Knee System1 de abril de 2026
Preguntas sobre este retiro
¿HCS - Headless Compression Screw System, Headless Compression Screw, Cannulated; Article Numbers: 31207-10, 31207-12, 31207-14, 31207-16, 31207-18, 31207-20, 31207-24, 31207-26, 31207-28, 31207-30, 31257-12, 31257-14, 31257-16, 31257-18, 31257-20, 31257-22, 31257-24, 31257-28, 31257-30, 31307-18, 31307-32, 31307-34, 31307-36, 31307-38, 31358-30, 31437-30, 31437-38, 31437-40, 31437-50, 31757-100, 31757-120, 31757-45, 31757-80, 31757-85, 31757-90. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1606-2026 el 1 de abril de 2026. Peligro reportado: Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · E-Mail