FDAFDA · Alimentos y medicamentos·N.º oficial Z-1606-2026
HCS - Headless Compression Screw System, Headless Compression Screw, Cannulated; Article Numbers: 31207-10, 31207-12, 31207-14, 31207-16, 31207-18, 31207-20, 31207-24, 31207-26, 31207-28, 31207-30, 31257-12, 31257-14, 31257-16, 31257-18, 31257-20, 31257-22, 31257-24, 31257-28, 31257-30, 31307-18, 31307-32, 31307-34, 31307-36, 31307-38, 31358-30, 31437-30, 31437-38, 31437-40, 31437-50, 31757-100, 31757-120, 31757-45, 31757-80, 31757-85, 31757-90.
Publicado el 1 de abril de 2026 · Inicio del retiro: 23 de enero de 2026
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Qué hacer
Voluntary: Firm initiated · E-Mail
Detalles del retiro
- Fabricante
- I.T.S. GmbH
- Distribuidor
- US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
- Lote
- Article Number (Lot Numbers): 31207-10 (679/112002), 31207-12 (679/092133, 679/102343, 679/112006), 31207-14 (679/102050), 31207-16 (679/102345, 679/102049), 31207-18 (679/0323110, 679/102355, 679/112001), 31207-20 (679/0323111, 679/102051, 679/102356), 31207-24 (679/102028), 31207-26 (679/112008), 31207-28 (679/112009), 31207-30 (679/0323140, 86/483377), 31257-12 (679/081904), 31257-14 (679/042323, 679/081905), 31257-16 (679/042324), 31257-18 (679/112308, 679/022128), 31257-20 (679/022123, 679/042326, 679/112321), 31257-22 (679/022124, 679/042327, 679/112325), 31257-24 (679/022125, 679/112326), 31257-28 (679/022130, 679/052306), 31257-30 (679/022131), 31307-18 (679/061907), 31307-32 (679/062285), 31307-34 (679/062286), 31307-36 (679/062287), 31307-38 (679/032118), 31358-30 (26/F67790), 31437-30 (679/071916), 31437-38 (679/071918), 31437-40 (679/032406, 679/071903), 31437-50 (679/071920), 31757-100 (35/51892), 31757-120 (86/320792), 31757-45 (86/320793), 31757-80 (26/F55872), 31757-85 (26/F55873), 31757-90 (26/F55874).
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- Austria
¿Vendes productos como este?